3 Technical data
11
F-100 / F-105
Operation Manual, Version C
3
Technical data
This section introduces the reader to the Recirculating Chiller and its specifications. It contains the
technical data, requirements and performance data.
3 .1
Technical data
Technical data
F-100
F-105
Dimensions (W x H x D)
280 x 500 x 420 mm
280 x 500 x 420 mm
Weight
28 kg
30 kg
Power consumption (max.)
850 W
850 W
Supply voltage
230 VAC ±10 %
115 VAC ± 10 %
230 VAC ±10 %
115 VAC ± 10 %
Fuse
230 V
115 V
6.3 AT
8 AT
6.3 AT
8 AT
Frequency
50/60 Hz
50/60 Hz
Environmental conditions
For indoor use only
- Temperature
5–35 °C
5–35 °C
- Altitude
up to 2000 m
- Humidity
Maximum relative humidity 80 % for temperatures up to 31 °C
decreasing linearly to 50 % relative humidity at 40 °C
Temperature display
No display
digital, resolution 0.1 °C
Overvoltage category
II
II
Degree of protection
IP20
IP20
Pollution degree
2
2
Cooling capacity at 15 °C
---
530 W
Cooling capacity at 10 °C
300 W
390 W
Cooling capacity at 0 °C
---
120 W
Cooling capacity at -10 °C
---
10 W
Refrigerant
R134a
R134a
Set temperature range
fix +10 °C
–10 °C … +25 °C
Temperature regulation accuracy ±2 °C
±1 °C
Tank volume
3 L
3 L
Tubing connection
8 mm (GL14)
8 mm (GL14)
Pump flow rate
2.5 L/min
2.5 L/min
Pump pressure (max.)
0.6 bar
0.6 bar
Содержание F-100