8 Troubleshooting
25
F-100 / F-105
Operation Manual, Version C
8
Troubleshooting
This section helps to resume operation after a problem has occurred with the instrument which does
not require special technical training. It lists possible occurrences, their probable causes and suggests
how to remedy the problem.
8 .1
Error message display (only F-105)
2
1
1
For every error the triangle lights.
2
The error number is indicated on the
display.
8 .2
Malfunctions and their remedies
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Remedy”.
Malfunctions and their remedies
Error Code
Problem
Remedy
E01
Temperature fault
Temperature sensor circuit break
Switch off the unit, and restart.
Call service if the problem persists.
E04
Pressure fault of the compressor
Switch off the unit, let it cool down the compressor.
Call service to check the system if the problem persists.
E05
Data fault
Switch off the unit and restart.
Call service if the problem persists.
E06
Temperature fault in the electronics
Switch off the unit, let it cool down, clean air inlet and restart.
Call service if the problem persists.
Problem
Remedy
F-100 / F-105 does not work Switch off the unit and unplug the
power cord.
Change the fuse according to technical
data and restart.
Call service if the problem persists.
1
1
Fuse