Melting Point M-560 Operation Manual, Version G
6.4.4
User calibration
A user calibration can be performed with any four substances at any heating ramp. Therefore the
procedure described in 6.4.2 can be performed with the user calibration mode.
6.4.5
Verification
To verify the calibration, proceed as follows:
• Measure all standards (BUCHI recommends using the verification set (11055019) but it is also
possible to use your internal standards.
NOTE
The verification is not a guided process.
• When all measuring results were within the required tolerances, click on
Verified
. The following
screen appears:
• Enter the following password:
VER
. The current date is now indicated under “Last verification”.
6.5
Settings, SysInfo, Test
Test:
Functional test of the instrument. To
perform the test of several functions follow
carefully the on-screen instuctions. If a function
test does not show
ok
, please contact the cus-
tomer service. The Test protocol is described in
section 6.5.1.
SysInfo:
Instrument gives information regarding
settings and connected devices. The SysInfo
protocol is described in section 6.5.2.
Settings:
This menu can be used to change
parameters.