6 Operation
22
Melting Point M-560 Operation Manual, Version G
6.2.2
Determination without pre-registered melting point method
• Choose the menu for melting point determination.
• Make sure that all samples are removed.
• Press START to begin the determination process immediately with the last parameters used. They
always remain saved in the instrument and are indicated in the green field in the lower part of the
screen.
NOTE
If the current parameters do not meet your requirements, press Edit and set start temperature, stop
temperature, and temperature gradient.
a b
c
d
a
Start temperature
b
Stop temperature
c
Temperature gradient
d
Edit
• To adjust the desired melting parameters,
press
Edit
.
• Chose
start temperature
with the rotary knob, and
press
Next
.
• Chose
stop temperature
with the rotary knob, and
press
Next
.
• Chose
temperature
gradient
with the rotary
knob, and press
Save
.
NOTE
Insert samples only when the start temperature is reached (as soon as prompted).