30
18.Specifications
Test:
Glucose
Sample:
Whole Blood
Principle of the test method: A
mperometric,
FAD- glucose dehydrogenase
Test sites:
Fingertip, forearm
Measurement time:
5 seconds
Measurement Range: 1
.1~33.3mmol/L ( 20~600 mg/dL).
Batteries: 2
x AAA 1.5 V Alkaline 24A LR03
Operating Condition:
10~40
℃
(50-104
℉
)
20~80% RH (non-condensing)
Storage/Transportation Condition:
2~30
℃
(35.6-86
℉
), below80% RH
Width:
50 mm
Length:
100 mm
Thickness:
20 mm
Weight:
~72g
Memory: 500 measurement results with date and time
Auto power off after 2 minutes without action when a
strip is inserted, or 60 seconds if there is no strip inserted.
Traceability of glucose monitoring system: The accuracy of
Blood Glucose Monitoring System was
SuperCheck Plus
assessed by comparing blood results obtained by patients
with those obtained using a YSI Model 2300 Glucose
Analyzer, which is calibrated by YSI calibrator solution,
refer to NIST SRM (standard reference material) # 917A
Clinical Dextrose. The device has certified to meet the
following standard: IVD 98/79/EC, EN 61010-1,
EN 61010-2-101, EN 61326-1, EN 61326-2-6, ISO 15197
(P.S. Please refer to the package insert that came with your
test strips, and you will also find information on how the
system works and on the performance characteristic reference.)
Lancets and lancing device are medical devices under
the Directive 93/42/EEC.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands
Manufacturer (Lancets) :
SteriLance Medical (Suzhou) Inc.
No. 68 Litanghe Road, Xiangcheng,
Suzhou 215133, China