27

Problem

Possible Cause

Solution

Repeat the test after the 

meter and strip are placed in 

a warmer environment and 

allow the meter to warm up 

for a while before retesting.

Test again following the user 

guide for correct glucose 

measurement process. If you 

see HI again, please call your

 medical doctor for advice 

immediately. 

Repeat the test after the 

meter and strip are placed in 

a cooler environment and 

allow the meter to cool down 

for a period before retesting.

Please replace the two AAA 

batteries.

Please check the test strip to 

see if it is damaged or used.   

In either case, please discard 

the strip and repeat the test 

using a new strip.

Repeat the test with a new 

test strip. If E_6 continues 

to show up on screen, please 

call a customer service 

representative.

The meter can perform the 

glucose measuring without 

storing the test results into 

the meter's memory.  Please 

write down the test results 

into the data logger before 

you call a customer service 

representative.

Test again following the user

 guide for correct glucose 

measurement process. If you 

see LO again, please call your 

medical doctor for advice 

immediately. 

The meter is operating

in an ambient 

temperature below 

10 

 or 50

 which is

the lowest range of 

operating temperature.

(1) 

The meter displays 

an E_1 error 

message.                     

(2) 

The meter displays 

an E_2 error 

message.   

                  

The test result is 

lower than the 

measuring range 

of 1.1 mmol/L

 (20 mg/dL).

(3) The meter displays 

an E_3 error message. 

                    

(5)

The meter displays

 an E_5 error 

message.                     

          

(6)

The meter displays

an E_6 error 

message.         

(8) 

The meter displays

HI on screen.

(9) 

The meter displays

LO on the screen.

The meter is operating 

in an ambient 

temperature above 

40 

 or 104

 which is

 the cap of operating 

temperature.

The battery is too low 

to operate the meter.

1.If you apply the 

sample before blood 

drop icon appears on 

the screen.

2.The inserted test strip 

has been wet or used.

The meter is not 

working properly, 

either because of a 

defective meter or 

a defective strip.

The memory chip 

of the meter could 

be damaged or 

malfunctioning.

The test result is 

higher than the 

measuring range 

of 33.3 mmol/L

 (600 mg/dL).

(4)

The meter displays 

an E_4 error 

message.     

Please re-test by inserting a 

new test strip ensuring an 

adequate amount of blood 

comes into contact with the 

test strip.

(7)

The meter displays

an E_7 error 

message.         

The blood sample is 

not sufficient .

Содержание 5228B

Страница 1: ...5228BT M01A1 Rev date 11 2013 USER GUIDE Model 5228B...

Страница 2: ...r to safety related notes in the manual accompanying this instrument Consult instructions for use Keep away from sunlight Authorised representive in the european community Manufacturer Do not re use U...

Страница 3: ...Blood Glucose Test 11 Recalling the Memory and Viewing the Average 12 Expected Values 13 Transmitting Results Maintenance 14 Maintenance of Your System 15 Troubleshooting 16 Customer Service 17 Limita...

Страница 4: ...us Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus nor for use with neonates The alternative site testing forearm in this system can only be use...

Страница 5: ...ancets 8 Lancing Device 9 Package Inserts for Test Strips Control Solution and Lancets Lancing Device A single level control solution Medium is provided in the Starter Kit and no control solution is i...

Страница 6: ...2 LCD Display Guide you through the test using symbols and simple messages 3 M Key Power ON OFF also for memory recalling mode please refer to manual for detailed function description 4 C Key Setting...

Страница 7: ...ip When to use AST Medication stress illness food and exercise can affect blood glucose levels Capillary whole blood at the fingertip can reflect test changes faster than capillary blood at other site...

Страница 8: ...or you to get a drop of blood for AST 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective top of the lancet 4 Close the end cap of the lancing device and grasp the lancing d...

Страница 9: ...of the vial and quickly recap the vial to keep air out 8 Insert the test strip into the meter while the meter is off The meter will turn on automatically A blinking blood icon will be displayed on the...

Страница 10: ...tting sufficient blood for a test it may be helpful to rub the puncture site a little longer before using the lancing device 11 Apply your blood to the front edge the tip of the test strip and the met...

Страница 11: ...icapped persons near you when you conduct a glucose test using this system Please keep the test strip vial away from children The test strips and vial cap can present a choking hazard Please be cautio...

Страница 12: ...inger and let a blood drop form 3 Apply the blood drop to the front edge of the test strip when the blood drop icon is still blinking on the meter Wait for 5 seconds and the meter will display the tes...

Страница 13: ...er in the opposite direction of the arrow to close the cover into position 11 Note 1 It is suggested that batteries need to be replaced when the battery icon appears empty If you keep using the meter...

Страница 14: ...ution test flag c 5 Indicates the meter is ready to take the blood sample when it flashes 6 Indicates the battery status 7 Indicates current displayed result is an average 8 Appears when alarm is on 9...

Страница 15: ...2029 After the year setting is confirmed you will see the month segment flashing on the screen Set the Month Press C key to adjust the month until the desired month is displayed and then press the M k...

Страница 16: ...clock in the lower left corner with a time segment display it means that you are in the alarm setting mode If you see only AL without the icon of the small alarm clock it means that you are not in the...

Страница 17: ...mmol L is the standard unit of measurement in the Europe If the meter is to be sold in the Europe the unit of measurement will be set to mmol L by default Delete Memory Press C key to toggle between a...

Страница 18: ...esults obtained during the 7 14 28 60 and 90 calendar days preceding the current date and time settings Note Your meter can show results either in mg dL or mmol L units It is very important to use the...

Страница 19: ...of screen indicating the control test mode The test result will not be stored in the memory 3 Squeeze a small amount of Control Solution on a flat surface and apply to the front edge of the test strip...

Страница 20: ...vial discard the first drop of control solution and wipe off the dispenser tip to ensure a good sample and an accurate result 6 Use only for 3 months after first opening Record the open date on the C...

Страница 21: ...he end cap counter clockwise 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective tip of the lancet 4 Close the end cap of the lancing device as shown Slide into locking posi...

Страница 22: ...automatically A blinking blood icon will be displayed on the screen If no action is taken in 2 minutes the meter will turn off automatically 8 Use the adjustable lancing device or the disposable lanc...

Страница 23: ...nd or by pushing the ejector button The meter will turn off and the test result is stored auto matically Important 1 Never reuse a test strip or a lancet An E_5 error message will be displayed in the...

Страница 24: ...f press M key to turn on the meter After a brief full screen display the meter will display the total number of test data stored in memory Then the screen will display 001 as the sequential number of...

Страница 25: ...rward from 7 day average to 90 day average The average display will move from 90 days backward to 7 days if you continue to press C key 3 Press the M key and hold for only 2 seconds to exit the recall...

Страница 26: ...ues 1 Reference SuperCheck Plus meter gives plasma equivalent results Before eating 5 6 mmol L 100mg dL Two hours after meals 7 8 mmol L 140mg dL Blood glucose levels normally will vary from time to t...

Страница 27: ...he meter has a shelf life of 2 years 14 Maintenance of your system We suggest you should periodically compare the test system to another test system which is well maintained and monitored by a healthc...

Страница 28: ...ple is not sufficient HI The glucose level is too high LO The glucose level is too low Following is a summary of all Error Messages These messages help to identify certain problems but do not appear i...

Страница 29: ...lease call your medical doctor for advice immediately The meter is operating in an ambient temperature below 10 or 50 which is the lowest range of operating temperature 1 The meter displays an E_1 err...

Страница 30: ...please contact Biotest Medical Corporation No 3 2 Chien kuo Road TEPZ Tantzu Taichung 427 Taiwan R O C TEL 886 4 2532 6668 E mail info mail biotestsystems com For questions related to your health con...

Страница 31: ...ructose observed in blood glucose test 7 Use only Heparin or EDTA for anticoagulation of fresh capillary or venous whole blood 8 Xylose Do not test blood glucose during or soon after xylose absorption...

Страница 32: ...ccuracy of Blood Glucose Monitoring System was SuperCheck Plus assessed by comparing blood results obtained by patients with those obtained using a YSI Model 2300 Glucose Analyzer which is calibrated...

Страница 33: ...edical Corporation No 3 2 Chien kuo Rd TEPZ Tantzu Taichung 427 Taiwan R O C E mail info mail biotestsystems com Reference 1 American Diabetes Association Diabetes Care January 2007 volume 30 Suppl 1...

Страница 34: ...and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help FCC Caution Any...

Страница 35: ...BiotesT Medical Corporation No 3 2 Chien kuo Road TEPZ Tantzu 427 Taichung Taiwan ROC Tel 886 4 25326668 Fax 886 4 25326528 MEDNET GmbH Borkstrasse 10 48163 M nster Germany 0537...

Отзывы: