9.Running a Control Solution Test
IMPORTANT:
Always make sure you press C key for Control Solution test,
otherwise the control solution test result will be stored in the
memory, and affect your AVG results
.
1
2
3
4
5
2.Press the C key, and the meter will
display a letter “C” on the upper right
corner of screen, indicating the control
test mode. The test result will not be
stored in the memory.
3.Squeeze a small amount of Control
Solution on a flat surface and apply to
the front edge of the test strip.
4.Wait for 5 seconds, and the meter will
display the result.
5.Remove the used strips by hand or by
pushing the ejector and the meter will
power off.
1. Insert a test strip into the meter while
the meter is off. The meter will turn on
automatically. A blinking blood icon
will be displayed on the screen.