1 ACCESS TO METROLOGICAL PARAMETERS
To access customisation of the metrological parameters (parameters more significantly linked to
legal operation of the scale) you must press the calibration button
1.1 Version with connectors
Figure 1.1-1 - Calibration button - Plastic terminal version
Figure 1.1-2 - Calibration button - Stainless steel terminal version
Key:
1. Stamped self-adhesive label sealing the access hole to the calibration switch
2. Calibration switch
Referring to the previous figures, proceed as follows:
Switch the terminal on.
Remove the self-adhesive label (ref. 1) which seals the access hole to the calibration switch.
Using a non-metal tool, press the calibration switch (ref. 2).
ATTENTION
Removing the metrological stamps makes it impossible to use the weighing terminal for
commercial transactions towards third parties. To restore them, contact the After-Sales Service of
the Assistance Network.
2
1
2
1