20
DC Voltage
Function RANGE
Accuracy
60.00mV
0.12%+2d
mV
600.0mV
0.06%+2d
V
9.999V, 99.99V,
999.9V
0.08%+2d
Input Impedance: 10M
, 50pF nominal
(80pF nominal for 600mV range)
AC Voltage
Function RANGE
Accuracy
50Hz ~ 60Hz
mV
60.00mV, 600.0mV
V
9.999V, 99.99V,
999.9V
0.5% + 3d
40Hz ~ 500Hz
mV
60.00mV, 600.0mV 0.8% + 4d
9.999V, 99.99V
1.0%+4d
V
999.9V
2.0%+4d
500Hz ~ 1kHz
mV
60.00mV, 600.0mV 2.0% + 3d
9.999V, 99.99V
1.0%+4d
V
999.9V
2.0%+4d
1kHz ~ 3kHz
mV
60.00mV, 600.0mV 2.0%+3d
V
9.999V, 99.99V,
999.9V
3.0%+4d
3kHz ~ 20kHz
mV
60.00mV
1)
,
600.0mV
1)
2%+3d
9.999V
2)
, 99.99V
3dB
V
999.9V
Unspec’d
1)
Specificied from 30% to 100% of range.
2)
for 3kHz ~ 15kHz only
Input Impedance: 10M
, 50pF nominal
(80pF nominal for mV ranges)
dBm
At 600
, -11.76dBm to 54.25dBm,
Accuracy:
0.25dB + 2d (@40Hz -- 20kHz)
Input Impedance: 10M
, 50pF nominal
Selectable reference impedance of 4, 8,
16, 32, 50, 75, 93, 110, 125, 135, 150, 200,
250, 300, 500, 600, 800, 900, 1000, 1200
AutoCheck
TM
(DCV)
RANGE
Accuracy
9.999V, 99.99V, 999.9V
0.5%+3d
Lo-Z DCV Threshold:
> +1.5VDC or < -1.0VDC nominal
Lo-Z DCV Input Impedance:
Initially approx. 3.0k
, 165pF nominal;
Impedance increases abruptly within a
fraction of a second as display voltage is
above 50V (typical). Ended up
impedances vs display voltages typically
are:
18k
@100V
125k
@ 300V
320k
@ 600V
500k
@ 1000V
DC
AC
& AC+DC
AC
Voltage
1)
Function
RANGE
Accuracy
DC, 50Hz ~ 60Hz
mV
60.00mV, 600.0mV
V
9.999V, 99.99V,
999.9V
0.7% + 6d
40Hz ~ 1kHz
mV
60.00mV, 600.0mV 1.0%+6d
V
9.999V, 99.99V,
999.9V
2.2%+6d
1kHz ~ 20kHz
mV
60.00mV
2)
,
600.0mV
2)
2.2%+6d
9.999V
3)
, 99.99V
3dB
V
999.9V
Unspec’d
1)
Specificied from 10% to 100% of range for
Model 821
2)
Specificied from 30% to 100% of range.