Operation
38 / 72
E164809_2_002
Experiment parameters that can be edited through the Acquisition Parameters menu:
Figure 6.10: The Acquisition Parameter Table
Scans (1 – 32000)
Scans is used to set the number of times the NMR experiment should be repeated for signal
averaging before the result is evaluated. The required number of scans depends on the
signal strength, the desired signal-to-noise (which will limit the precision of the
measurement), and the desired speed of analysis (time between scans is set by the Recycle
Delay). Signal to noise improves with the square root of the number of scans. The remaining
number of scans during an experiment is shown in the lower right corner of the display.
RD (seconds)
RD, or
R
ecycle
D
elay, is used to set the time delay in seconds between repeated scans. The
recycle delay is usually set just a little longer than the relaxation time of the sample. If the
delay is deliberately made shorter than the sample relaxation time, one or more dummy shots
can be made prior to the experiments in order to ensure that the first scan has the same
history as all subsequent scans.
Gain (mq-Series NF systems: 40 – 119dB)
Gain sets the signal amplification in the receiver. It is a common procedure to set the gain
such that the highest part of the signal reaches about 80% of the signal input range. It is
known that the noise start to increase significantly when exceeding 95…100dB. As the NF
minispec features a high signal resolution like 14/15 bit, it is also possible to work with lower
input signal levels, thus avoiding gain values that are too high.
Note that the gain should not be changed for calibrated applications. If the range of input
signals is too large for one gain setting, it may be better to have several calibrations with
different gain settings.
Содержание minispec LF Series
Страница 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...
Страница 8: ...Introduction 8 72 E164809_2_002...
Страница 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...
Страница 14: ...Safety 14 72 E164809_2_002...
Страница 18: ...Transport Packaging and Storage 18 72 E164809_2_002...
Страница 50: ...Operation 50 72 E164809_2_002...
Страница 56: ...Maintenance 56 72 E164809_2_002...
Страница 58: ...Replacement of Parts 58 72 E164809_2_002...
Страница 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...
Страница 66: ...Contact 66 72 E164809_2_002...
Страница 68: ...List of Figures 68 72 E164809_2_002...
Страница 71: ...E164809_2_002 71 72...
Страница 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...