Operation

26 / 72

E164809_2_002

Normally the tables are executed after 

RUN

 is selected, beginning from row one to the end of

the  entries.  But  you  can  also  measure  single  rows  of  the  table  by  marking  these  lines  and
then  selecting 

RUN

.  This  may  be  useful  if  not  all  samples  are  prepared  at  the  beginning  of

the analysis or if several measurements should be repeated.

Spreadsheet data are saved automatically every time an entry is changed and any time the
ENTER key is pressed. The filename is generated by the minispec software:

<application>\MMMYY\DD\hh\mm\ss.mdb

 with:

<application>

: path of the current application

MMM

: month (Jan, Feb, Mar,..., Nov, Dec)

JJ

: year (99, 00, 01, ...)

DD

: day (01, 02, ..., 31)

hh

: hour (00, 01, ...23)

mm

: minute (00, 01, ..., 59)

ss

: second (00, 01, ..., 59)

Saving in ASCII-format

The  files  can  be  saved  alternatively  in  ASCII-format  by  using  the 

Save

  or 

Save  Copy  As

commands.  The  file  name  will  be  prompted  in  a  standard  file  selector  box,  and  the  file
extension used is *

.tab

. The first row of the data includes the headlines of the spreadsheet,

the columns are spaced with a <TAB>. The ASCII tables can also be opened with standard
text editors like NOTEPAD/WORDPAD or MS-EXCEL.

During  a  measurement  the  spreadsheet  is  updated  by  the  minispec  software  every  time  a
result  is  added.  The  tables  can  be  opened  on  the  measurement  PC  or  any  other  PC
equipped  with  the  minispec.exe  software  or  MS-ACCESS  2013.  External  monitoring  of  the
measurement results via a network connection is therefore possible.

6.2.4

minispec.exe Menu Functions

The  minispec  software  provides  many  opportunities  for  data  display,  data  saving,  and
measurement  handling.  This  chapter  describes  all  functions  that  are  available  when  the
software is configured as 

NMR-Analyzer

.

When  the  minispec  is  used  only  for  specialized  applications  (Fluorine,  etc.)  many  of  these
functions  are  unnecessary.  In  this  case,  the  number  of  menu  options  is  reduced  to  simplify
the  interface  and  make  the  software  easier  to  use  when  the  configuration  is  adjusted  to
certain  analyzer  types.  Commands  whose  functionality  may  be  not  reasonable  in  some
contexts  appear  grayed  or  may  no  longer  be  visible.  The  actions  of  the  functions  are
described below.

6.2.4.1

The Bars

The  minispec  software  menu  structure  and  toolbars  are  designed  within  the  framework  of
most  standard  Windows  software  to  ensure  intuitive  handling  (see  the  screen  layout  in  the
figure below). In the 

Menu 

bar actions can be selected via pull-down menus. Icons are also

used  for  commonly  used  actions  and  remain  visible  in  the 

Standard 

toolbar.  Depending  on

the  options  selected,  the  most  important  commands  for  the  execution  and  abortion  of
measurements  may  also  be  accessible  through  large  icons  on  a 

Measurement 

toolbar,

which  is  usually  positioned  at  the  bottom  of  the  screen  below  the  signal  window.  In  the

Result 

bar  the  numerical  result(s)  of  the  last  measurement  is  displayed.  The 

Info 

toolbar

contains easily selectable buttons pertaining to the instrument status and parameter display,
as  well  as  a  list  of  loaded  applications.  Finally,  the  status  of  the  measurement  and
connections are summarized in the 

Status

 bar. The functionality of these items is described

in the following sections in detail.

Содержание minispec LF Series

Страница 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...

Страница 2: ...reproduced stored in a retrieval system or transmitted in any form or by any means without the prior consent of the publisher Product names used are trademarks or registered trademarks of their re sp...

Страница 3: ...ription 19 4 2 Unit Description 19 4 3 Movable Working Table Cart 20 5 Installation 21 5 1 Siting Considerations 21 5 1 1 Environmental Considerations 21 6 Operation 23 6 1 Turning the Unit On 23 6 2...

Страница 4: ...53 8 1 Scans 53 8 2 Recycle Delay Time 53 8 3 Gain 53 8 4 Acquisition Pre Delay Time 54 8 5 Dummy Shots 54 9 Maintenance 55 9 1 Cleaning 55 10 Replacement of Parts 57 10 1 Returning the Unit for Repai...

Страница 5: ...ies will benefit from the software introduction in the section minispec Software GUI To provide a better understanding of the principles involved an introduction to the basics of NMR is provided in se...

Страница 6: ...entation for Bruker Service technicians Before starting any work personnel must read the manual thoroughly and understand its contents Compliance with all specified safety and operating instructions a...

Страница 7: ...me Every effort has been made to avoid errors in text and figure presentation in this publication In order to produce useful and appropriate documentation we welcome your comments on this publication...

Страница 8: ...Introduction 8 72 E164809_2_002...

Страница 9: ...ly be set up and operated in conjunction with the instructions in this manual Only qualified personnel should be allowed to work on this equipment The primary installation maintenance and repair of th...

Страница 10: ...nor injuries or material damage but may also result in a life threatening situation if hazardous substances are used u If glassware or ceramics breaks refer to the corresponding precautions and cleani...

Страница 11: ...se the power cable that is delivered with the product If replacement of the power cable is necessary make sure that the ratings with regard to the power consumption of the device are suitable WARNING...

Страница 12: ...ks and electronic devices may be damaged u Mark the magnetic field of more than 0 5 mT 5 Gauss before start up u Keep people with active medical implants or heart pacemakers away from the 0 5 mT 5 Gau...

Страница 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...

Страница 14: ...Safety 14 72 E164809_2_002...

Страница 15: ...he content may be damaged Fragile Marks packages with fragile or sensitive contents Handle the package with care do not allow the package to fall and do not allow it to be impacted Protect Against Moi...

Страница 16: ...diately inspect the delivery for completeness and transport damage Proceed as follows in the event of externally apparent transport damage Do not accept the delivery or only accept it subject to reser...

Страница 17: ...he packages under the following conditions Do not store outdoors Store in dry and dust free conditions Do not expose to aggressive media Protect against direct sunlight Avoid mechanical shocks Storage...

Страница 18: ...Transport Packaging and Storage 18 72 E164809_2_002...

Страница 19: ...all default settings and the user documentation 4 1 Brief Description All systems of the minispec LF product range have an integrated basic probe module where the respective sample is positioned usin...

Страница 20: ...table operation the system is placed and slip resistant screwed on a movable working table or so called cart The cart is standard with the shipment of all minispec LF systems As special feature There...

Страница 21: ...objects or within 2 m of moving metallic objects CAUTION Accident or material damage hazard from falling or tilting instrument Instrument may fall down or tilt during installation retrofitting disman...

Страница 22: ...damage to health Ferromagnetic tools in the magnetic field are significantly hazardous Disks and electronic devices may be damaged u Mark the magnetic field of more than 0 5 mT 5 Gauss before start u...

Страница 23: ...ramming differs slightly for both systems For minispec Plus users the minispec exe will only be used in a few exceptional cases including Checking the connection between PC and minispec system in some...

Страница 24: ...e startup To add applications to the list use Browse to access a file selector box or use Delete to remove applications from the list To choose another minispec use the Select button in the Select min...

Страница 25: ...opy As from the pulldown menu File 6 2 3 2 The Results Window The Result Window is an output device for minispec applications that uses the ExpSpel command print_line RESULTBOX to display a record of...

Страница 26: ...u Functions The minispec software provides many opportunities for data display data saving and measurement handling This chapter describes all functions that are available when the software is configu...

Страница 27: ...e File Menu The list of items in the Menu Bar depends on the active window If the Signal Window is active and an application is loaded all entries are visible The functions of the menu can be activate...

Страница 28: ...his command to close all windows in the active application or document If the focus is on one of the four views windows of a minispec application the complete minispec application with all views will...

Страница 29: ...Prompt Saving to deactivate the prompt when saving in which case the unsaved parts of the result box are then saved automatically After reloading the minispec application the results will be erased b...

Страница 30: ...view toolbar offers options to view either one or two pages at a time movement back and forth through the document zoom in and out of pages and to initiate a print job The Print Setup Command Use this...

Страница 31: ...data in ASCII format to allow an easy transfer to Excel Use Copy as Bitmap from the Edit menu to copy signals as bitmap to the clipboard Shortcut Keys CTRL B Copy as Bitmap Shortcut Keys CTRL C Copy...

Страница 32: ...d Once the system requirements for TCP IP have been fulfilled a software link between the minispec software running on a PC and the minispec must be initialized To establish the PC to minispec connect...

Страница 33: ...connected minispec customize your minispec software through the Welcome Box to access the calibration parameters The Advanced button in the settings table allows access to a list of instrument setting...

Страница 34: ...calibrated minispec instrument is indicated by a yellow mark Check To run the Daily Check use the corresponding toolbar button or select Daily Check from the menu If one of the checks fails it indicat...

Страница 35: ...uccessful license procedure the menu entry is checked and grayed Have ready the minispec instrument s system number see Status Bar and contact BRUKER for information or license numbers 6 2 4 5 The Mea...

Страница 36: ...r fat content from the NMR signal As a consequence the NMR signal must first be calibrated with respect to sample properties Select Calibrate Sample from the Measurement menu Shortcuts Standard Tool B...

Страница 37: ...ccessive repetitions and the maximum number of repetitive measurements 6 2 4 6 The Parameter Menu All parameters for acquisitions applications and repetitive measurements can be defined with the assis...

Страница 38: ...The recycle delay is usually set just a little longer than the relaxation time of the sample If the delay is deliberately made shorter than the sample relaxation time one or more dummy shots can be ma...

Страница 39: ...F systems come with a real digital filter which means that the dwell time is related to the setting of the digital filter Changing the digital filter from the standard value in the application softwar...

Страница 40: ...figured as the Relaxation Time or NMR Analyzer type all toolbars are docking toolbars This means that the toolbars can be moved to any location on the entire screen by pressing and holding the left mo...

Страница 41: ...rement results The Statistics Box contains Mean value Standard deviation Current data value Total number of data Maximum value Minimum value Difference between minimum and maximum value Graphic repres...

Страница 42: ...s between separate or overlapping displays of real and imaginary signals Toggles between Full Scale the limits are the highest and lowest values to be displayed and normal 100 signal scale Zoom Scale...

Страница 43: ...Diagnostics View Command With this command a window is opened up which displays various information for diagnostic purposes These options are helpful for checking the instrument status connections and...

Страница 44: ...This command is only available in connection with calibrations You can fit your calibration data and display the fit results see Info Views point 4 The FFT Magnitude Power and Absolute Commands A Fas...

Страница 45: ...s at the bottom of the main window If there is an open document window at the bottom of the main window some or all of the icons may not be visible because they are hidden behind this document window...

Страница 46: ...ice and version number of your copy of the minispec software Shortcuts Standard Button Bar Additional Help Functions While moving the cursor through the buttons a short on line help will be displayed...

Страница 47: ...Figure 6 17 minispec Plus NF Icon The software will start and come to the log in screen Figure 6 18 The minispec Plus Log in Dialog To get full access to all the menus of minispec Plus type ADMIN and...

Страница 48: ...nstrument should be valid for 24 hours In case there is a Daily Check failure it is recommended to run the Update Settings procedure and to follow the corresponding instructions In any case if the Upd...

Страница 49: ...ec you want to use and configure the software e g choose applications or an analyzer type 3 Depending on the application you use see your application description start your measurement or calibration...

Страница 50: ...Operation 50 72 E164809_2_002...

Страница 51: ...protons 1 H have a resonance frequency of 20 MHz The orientation of the magnetic field generated by the RF pulse is perpendicular to the static external polarizing field When for example a sample con...

Страница 52: ...analysis For example because of their different relaxation behavior solid and liquid parts in one sample can be distinguished clearly and their amount determined as in Solid Fat Content SFC measureme...

Страница 53: ...set by the Recycle Delay Signal to noise improves with the square root of the number of scans The remaining number of scans during an experiment is shown in the lower right corner of the display 8 2...

Страница 54: ...low the temperature of the sample to equilibrate in the magnet The Acquisition Pre Delay Time option allows such a delay to be set in seconds The waiting time is added to the recycle delay of the firs...

Страница 55: ...the Main Unit Use a clean damp cloth to wipe off any residue from the outside of the unit Wipe off excess moisture immediately with a dry cloth For non water soluble marks wipe using a cloth with a fe...

Страница 56: ...Maintenance 56 72 E164809_2_002...

Страница 57: ...ssary information for the subsequent repair process steps 2 They will provide you with details on the shipping address and also in most cases a Return Merchandise Authorization number RMA number that...

Страница 58: ...Replacement of Parts 58 72 E164809_2_002...

Страница 59: ...nvironmental protection regulations 11 2 Disposal in Europe Environmental information for laboratory and industrial customers within the EU European Union This laboratory product is developed and mark...

Страница 60: ...ffers to take back the equipment only for deliveries after 23 03 2006 after termination of use at the customer site upon request by the customer Additional costs for dismantling and transport service...

Страница 61: ...plate in the bottom which can be used to position the control unit The frame is included in the specifications The system comes on a cart and is included in the specifications 12 2 Connection Values E...

Страница 62: ...3 4 CAN High 5 5V DC maximum 150 mA control board 6 7 5V DC maximum 150 mA control board 8 GND Lamda power supply 9 10 11 CAN Low 12 GND control board 13 14 5V DC maximum 150 mA control board 15 NC T...

Страница 63: ...condensing 50 IP20 Table 12 5 Operating Environment NormaI environmental conditions CAN CSA 61010 1 12 IEC 61010 1 2010 ANSI UL 61010 1 Indoor use only Maximum operation altitude 2000 m Ideal operati...

Страница 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...

Страница 65: ...s Contact our Bruker BioSpin service centers Bruker BioSpin provides dedicated hotlines and service centers so that our specialists can respond as quickly as possible to all your service requests appl...

Страница 66: ...Contact 66 72 E164809_2_002...

Страница 67: ...ndows 31 Figure 6 6 The minispec Pull Down Menu 32 Figure 6 7 The minispec Settings Table 33 Figure 6 8 The Measurement Pull Down Menu 35 Figure 6 9 The Parameter Pull Down Menu 37 Figure 10 The Acqui...

Страница 68: ...List of Figures 68 72 E164809_2_002...

Страница 69: ...l command 31 Exit Command 30 ExpSpel license number 24 F File menu 27 Find command 31 I I O Port Settings command 35 Inspect the delivery 16 Instrument Settings command 33 L Licenses command 35 List o...

Страница 70: ...72 E164809_2_002 T Technical Data 61 Toolbars 26 Transport and handling 15 U Unselect All command 31 Update Settings 23 Update Settings command 34 User interface 24 W Wake up configuration 23 Welcome...

Страница 71: ...E164809_2_002 71 72...

Страница 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...

Отзывы: