3-6
Basic Operations and Maintenance
USING THE MANUAL STAGE-POSITIONING CONTROLS
If your DektakXT system includes the manual sample-positioning stage (see
), you must
use the procedures described in this section to adjust both the X and Y positions.
Adjusting the X-Y Position
To position a sample using the manual sample-positioning stage controls:
1
Ensure the stylus is not touching the sample surface. If it is, click the
Tower Up
button
on the
toolbar.
2
Position the scan start site by using the X-Y positioning levers.
•
Pull the lever below the front of the platform for coarse X movement.
•
Pull the lever below the left side of the platform for coarse Y movement.
•
Use the respective knobs to finely position the sample.
Figure 3-8: X-Y Positioning Levers on the Manual Sample-Positioning Stage
NOTE –
After fine-tuning the scan site using the X and Y positioning levers, you may have
to re-null the tower to adjust for skew. (To re-null the tower, tower up and then tower down.)
Leveling the Stage
All stage configurations of the DektakXT system allow you to manually level the sample-positioning
stage by turning the leveling knob below it (see
). The closest possible manual leveling
ensures the best profiler performance. Guidelines for leveling the stage appear after the picture.
X coarse-
positioning Lever
Y coarse-
positioning
lever
X fine-positioning
knob
Содержание DektakXT
Страница 1: ......
Страница 2: ...P N 980 365 DEKTAKXT STYLUS PROFILER USER MANUAL ...
Страница 8: ...Table of Contents v Applying Filters 5 11 Performing Analyses 5 12 3D Mapping Optiion 5 13 ...
Страница 9: ......
Страница 25: ......
Страница 67: ......
Страница 71: ...5 4 Using Vision64 Figure 5 4 Measurement Setup Vision View Figure 5 5 Automation Setup Vision View ...
Страница 72: ...Using Vision64 5 5 Figure 5 6 Data Acquisiton Vision View Figure 5 7 Data Analysis Vision View ...
Страница 81: ......