Adjustment of a Chamber on the D8
40
DOC-M88-EXX268_V1_12.2016
Figure 4.3: Insert a deflection angle in CONFIGURATION for the temperature stage
1. Determine the ZI-value (figure ) and insert it as a deflection angle in
CONFIGURATION | Basic instrument information | Device | Goniometer | Center |
PAAR_Z_HTK1200
for the temperature stage (see figure
above).
2. Save the configuration and activate it.
3. Repeat steps
Performing a rocking curve
and
Rotation alignment by introducing
a software offset
to check if the deviation is within the range of 0° ± 0.02°.
4. Repeat step
Height check
for a final height check.
4.1.2
Alternative Procedure Without Alignment Slit
This method is less exact compared to the first one but more universal if no alignment slit
is available. It is important that the diffractometer is well aligned, i.e. tube/theta and detec-
tor in line at 0° position.
For correcting the chamber rotation, only the method for inserting a deflection angle in the
software is described here. If you want to rotate the chamber mechanically, follow the in-
structions below up to the point “Perform a rocking curve scan”, and continue with sub-
chapter
Adjustment Procedure with Alignment Slit [
alignment criterion specified in point 3 can then be relaxed to 0° ± 0.05°.
Содержание Anton Paar
Страница 6: ...Contents 6 DOC M88 EXX268_V1_12 2016 ...
Страница 12: ...List of Figures 12 DOC M88 EXX268_V1_12 2016 ...
Страница 26: ...General Remarks 26 DOC M88 EXX268_V1_12 2016 ...
Страница 34: ...Vacuum Equipment 34 DOC M88 EXX268_V1_12 2016 ...
Страница 58: ...Installation and Operation of the HTK Series TTK 600 XRK 900 and HPC 900 Chambers 58 DOC M88 EXX268_V1_12 2016 ...
Страница 111: ...Appendix Service Details DOC M88 EXX268_V1_12 2016 111 Figure 10 15 CHC_Stage Sample Temperature Details ...
Страница 135: ...DOC M88 EXX268_V1_12 2016 135 ...
Страница 136: ...Bruker Corporation info baxs bruker com www bruker com Order No DOC M88 EXX268 ...