50
ZTKS0172 / Z31964 / 04
Troubleshooting
Table continued
Indicator
Possible reason
Solution
Performed by
The pneumatic
isolator of the
magnet stand does
not reach the
operating position.
Pneumatic
controller in DOWN
position.
Switch pneumatic controller
to UP position.
Approved
Customer
Personnel
The pressure of the
gas supply is too
low.
Check the pressure of the
pneumatic supply. It must be
in the range of 5 to 8 bar
(70 to 112 psi).
Approved
Customer
Personnel
If the problem is still not
solved, contact Bruker
Service.
Approved
Customer
Personnel
The magnet
system is not level.
Stop the pneumatic
isolators. Check the level of
the cryostat.
Bruker Service
Defective leveling
valve.
Replace leveling valve or
isolator.
Bruker Service
Defective
membrane of an
isolator.
Replace leveling valve or
isolator.
Bruker Service
The movement of
the magnet stand is
not smooth.
Piston is not
centered or is
touching its casing.
Align magnet stand.
Bruker Service
Audible loss of gas.
Defective
membrane or
defective leveling
valve of an isolator.
Replace leveling valve or
isolator.
Bruker Service
Hose connector is
defective or loose.
Insert hoses correctly and
tighten screws.
Bruker Service
The movement of
the magnet stand is
too fast.
Incorrect adjust-
ment of the flow
control valve.
Close fully the flow control
valve; then open ½ turn.
Bruker Service
Содержание 500'89 Ascend Aeon
Страница 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Страница 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Страница 8: ...8 ZTKS0172 Z31964 04 Contact ...
Страница 28: ...28 ZTKS0172 Z31964 04 Safety ...
Страница 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Страница 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Страница 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Страница 81: ...81 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend 9 21 Current Lead Figure 9 12 Current lead 55 pin ...
Страница 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Страница 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Страница 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Страница 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Страница 98: ...98 ZTKS0172 Z31964 04 Index ...
Страница 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...