40
ZTKS0172 / Z31964 / 04
Troubleshooting
6.2
Problem
6.2.1
During Transportation
6.2.2
During Assembling
Indicator
Possible reason
Solution
Performed by
Tilt Watch / Shock
Watch activated.
Careless
transportation.
1.
Accept delivery with
reservation.
2.
Remark the extent of
damage in the trans-
portation documents.
3.
Start complaint process.
Approved
Customer
Personnel
Visible damage.
Careless
transportation.
1.
Accept delivery with
reservation.
2.
Remark the extent of
damage in the trans-
portation documents.
3.
Start complaint process.
Approved
Customer
Personnel
Indicator
Possible reason
Solution
Performed by
Ceiling height too
low for assembling
on magnet stand.
Site does not meet
the required
conditions.
Choose another site that
meets the required
conditions.
Bruker Service
Ceiling height too
low for inserting the
Helium Level
Sensor.
Site does not meet
the required
conditions.
Insert the Helium Level
Sensor before mounting the
magnet stand.
Bruker Service
Helium bore and
radiation shield are
not centered.
Alignment is not
correct.
Check fitting of the
alignment rods.
Bruker Service
Alignment rod is
loose or broken.
Replace alignment rod
a
.
Bruker Service
Reduction flange is
not centered.
Check the position of the
reduction flange.
Bruker Service
Содержание 500'89 Ascend Aeon
Страница 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Страница 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Страница 8: ...8 ZTKS0172 Z31964 04 Contact ...
Страница 28: ...28 ZTKS0172 Z31964 04 Safety ...
Страница 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Страница 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Страница 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Страница 81: ...81 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend 9 21 Current Lead Figure 9 12 Current lead 55 pin ...
Страница 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Страница 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Страница 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Страница 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Страница 98: ...98 ZTKS0172 Z31964 04 Index ...
Страница 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...