Display and operating elements
VIBROCONTROL 920
VC920E / C102833.002 / V 13
Page 17 / 33
5.4
Parameter: Group 2
Mode 1 Code number
Value: 22
Mode 2 Sensor sensitivity
Default value: 100 mV/mm/s
At the same time the sensitivity of an acceleration sensor is selected, the
integration for conversion of vibration acceleration to vibration velocity is
activated.
Selection
Value
Sensor for
Requirement
1
70 mV/mm/s
v
2
100 mV/mm/s
v
3
100 mV/g
a
-24 V
4
10 mV/g
a
+ 4 mA
5
100 mV/g
a
+ 4 mA
6
10 mV/g
a
-24 V
Meaning:
a = Vibration acceleration
v = Vibration velocity
Mode 3 Correction factor
Default value: 1.00
To allow sensors which have a sensitivity other than those listed in parameter
2 of Group 2 to be used with the instrument, the selected value can be
corrected with a factor in the range 0.1 to 1.99.
Mode 4 DC Current output range
Default value: 4 … 20 mA
Selection
Range
0
0 ... 20 mA
4
4 ... 20 mA