Chapter 3 — Operating Characteristics
Sensitivity
Preamplifiers Types 2663 and 2663 B
BE 1068 – 12
27
Fig.3.6
Frequency response of Type 2663 for different gain settings when operating beyond 33 kHz
3.3
Sensitivity
The preamplifiers are factory preset to have a unified output sensitivity of 1 mV/pC
(±l%). However, for operation with accelerometers of different charge sensitivity, this
may be readjusted (as discussed in
) for an output sensitivity between 1
and 100 mV/pC. For sensitivities greater than 25 mV/pC, the minimum lower limiting
frequency increases.
shows the minimum lower limiting frequency for given
gain settings.
Fig.3.7
Minimum lower limiting frequency versus gain setting
1 kHz
10 kHz
0 dB
2 dB
4 dB
6 dB
8 dB
10 dB
100 kHz 200 kHz
1 dB
3 dB
1 mV/pC
10 mV/pC
50 mV/pC
100 mV/pC
500 Hz
893398e
892362e
Preset value of 1 mV/pC
1.0
0.5
1.5
100
75
25
50
1
mV/pC
Hz
Содержание 2663
Страница 2: ...BE 1068 12 Preamplifiers Types 2663 and 2663 B User Manual ...
Страница 3: ......
Страница 8: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 1 Chapter 1 Introduction ...
Страница 12: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 5 Chapter 2 Operation ...
Страница 30: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 23 Chapter 3 Operating Characteristics ...
Страница 38: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 31 Chapter 4 Accessories ...
Страница 43: ...Preamplifiers Types 2663 and 2663 B User Manual Brüel Kjær 36 ...
Страница 44: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 37 Chapter 5 Specifications ...
Страница 46: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 39 Chapter 6 Service and Repair ...
Страница 48: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 41 Appendix A Ground Loop Noise ...
Страница 50: ...Preamplifiers Type 2663 and 2663 B User Manual BE 1068 12 43 Index ...
Страница 53: ......