Chapter 2 — Operation
Circuit Adjustment for Velocity Measurement
Preamplifiers Types 2663 and 2663 B
BE 1068 – 12
19
Select the frequency
F
LP
at which the integration of the signal is to begin.
Thus a value for the resistance
R
L
can be found in the following formula (2.6.1).
(2.61)
A suggested value for the frequency
F
LP
at which integration begins is 10 Hz since
this will suppress any low frequency noise caused by transducer/temperature varia-
tions.
Fig.2.13
Resistor connections when using the preamplifiers as an integrator
2.6.2
Changing the High-pass Filter
The two resistors which determine the characteristics of the high-pass filter are
and .
The value of resistor
is determined by the damping coefficient
and the cut-off
frequency
F
HP
of the filter and is determined by the following formula:
Select a value for
F
HP
=
F
LP
= 10 Hz.
R
L
1.59
10
8
F
LP
--------------------------
=
R
H
1
R
H
2
R
H
2
R
H
2
2
1.2
+
+
2
F
HP
C
---------------------------------
=
Содержание 2663
Страница 2: ...BE 1068 12 Preamplifiers Types 2663 and 2663 B User Manual ...
Страница 3: ......
Страница 8: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 1 Chapter 1 Introduction ...
Страница 12: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 5 Chapter 2 Operation ...
Страница 30: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 23 Chapter 3 Operating Characteristics ...
Страница 38: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 31 Chapter 4 Accessories ...
Страница 43: ...Preamplifiers Types 2663 and 2663 B User Manual Brüel Kjær 36 ...
Страница 44: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 37 Chapter 5 Specifications ...
Страница 46: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 39 Chapter 6 Service and Repair ...
Страница 48: ...Preamplifiers Types 2663 and 2663 B User Manual BE 1068 12 41 Appendix A Ground Loop Noise ...
Страница 50: ...Preamplifiers Type 2663 and 2663 B User Manual BE 1068 12 43 Index ...
Страница 53: ......