TECHNICAL CHARACTERISTICS
Non-contact infrared thermometer
Brother Max 6-in-1 Thermometer
Taking the body temperature at the forehead
• Room temperature working range: 10-45°C (55-113°F)
(1)
• Resolution: 0.1
• Measurement range: 34.0-42.5°C (93.2-108.5°F)
Accuracy °
C
from 34.0 to 35.9°C:
±0.3°C
from 36.0 to 39.0°C:
±0.2°C
(2)
from 39.1 to 42.5°C:
±0.3°C
Accuracy °
F
from 93.2 to 96.7°F:
±0.5°F
from 96.8 to 102.2°F:
±0.4°F
(2)
from 102.3 to 108.5F: ±0.5°F
Other readings
• Room temperature working range: 5-45°C (41-113°F)
(1)
• Resolution: 0.1
• Measurement range: 1.0-80.0°C (33.8-176°F)
Accuracy °
C
from 1.0 to 19.9°C:
±1.0°C
from 20.0 to 35.9°C:
±0.3°C
from 36.0 to 39.0°C
±0.2°C
from 39.1 to 42.5°C
±0.3°C
from 42.5 to 80°C:
±1.0°C
Accuracy °
F
from 33.8 to 67.9°F
±1.8°F
from 68.0 to 96.7°F
±0.5°F
from 96.8 to 102.2 °F
±0.4°F
from 102.3 to 108.5°F
±0.5°F
from 108.7 to 176 °F
±1.8°F
(1)
When the room temperature is between 10 and 15.9°C
(50 and 60.6ºF) and the “FACE” button is used, or when
the temperature is between 5 and 9.9°C (41 and 50ºF) and
the “HOME” button is used, accuracy and operating range
cannot be guaranteed and the display shows "Lo.5" and
the value alternately.
When the room temperature is between 40 and 45°C (104
and 113ºF), and the “FACE” or “HOME” buttons are
pressed, accuracy and operating range cannot be
guaranteed and the display shows "Hi.4" and the value
alternately.
(2)
The ASTM E1965-98-2009 standard re/-0.2°C
(+/-0.4°F) accuracy for infrared thermometers in the 37-
39°C (98.6-102.2°F) temperature range while the ASTM
E667-86 and E1112-86 standards r/- 0.1°C (+/-
0.2°F) precision for mercury and electronic thermometers in
the same temperature range.