74
MANUAL P/N 316005 REV: B
9.5 Theory of Operation
The Linear Analog Output Module circuit consists of an embedded microcontroller and associated circuitry.
The microcontroller picks off display data to the display module during the appropriate bus timing cycles and
performs a digital-to-analog conversion by using the external associated circuitry.
9.6 Troubleshooting
Symptom
Possible Cause
µP Failure LED is illuminated
or flashing
Microcontroller failure
There are no other user troubleshooting procedures associated with the Linear Analog Output Module. If you
are having difficulty with this module, contact Granville-Phillips Customer Service.
9.7 Specifications
Analog output loading: ± 2mA max.
Thermocouple Output
Pressure Decade
Range
Output Voltage
Smallest
Increment
Comment
< 10
-3
Torr
0 Vdc
-
Zero Vdc output below 1 mT
10
-3
Torr
10 - 99 mV
1 mV
1 mV = .1 mTorr
10
-2
Torr
.1 V - .99 V
10 mV
0 mV = 1 mTorr
10
-1
Torr
1.0 V - 9.9 V
100 mV
100 mV = 10 mTorr
t
1 Torr
10.0 V
-
Содержание Granville-Phillips Convectron 316 Series
Страница 11: ...12 MANUAL P N 316005 REV B Figure 1 3 Mounting Configurations...
Страница 26: ...MANUAL P N 316005 REV B 27 Figure 2 3 Convectron Gauge Indicated vs True Pressure Curve 0 1 to 100 mTorr...
Страница 27: ...28 MANUAL P N 316005 REV B Figure 2 4 Convectron Gauge Indicated vs True Pressure Curve 0 1 to 1000 Torr...
Страница 28: ...MANUAL P N 316005 REV B 29 Figure 2 5 Convectron Gauge Indicated vs True Pressure Curve 0 1 to 1000 Torr...
Страница 29: ...30 MANUAL P N 316005 REV B Figure 2 6 Convectron Gauge Indicated vs True Pressure Curve 10 4 to 0 1 mbar...
Страница 30: ...MANUAL P N 316005 REV B 31 Figure 2 7 Convectron Gauge Indicated vs True Pressure Curve 0 1 to 1000 mbar...
Страница 31: ...32 MANUAL P N 316005 REV B Figure 2 8 Convectron Gauge Indicated vs True Pressure Curve 0 1 to 1000 mbar...
Страница 75: ...76 MANUAL P N 316005 REV B...