MANUAL P/N 316005 REV: B
Example 4: True to indicated pressure conversion.
Assume you want to obtain a helium pressure of 100 Torr in the system. On Figure 2-4, locate 100 Torr on
the left hand scale, travel horizontally to the right to attempt to intersect the He curve. Because the
intersection is off scale it is apparent that this true pressure measurement requirement for helium exceeds
the capability of the instrument.
For gases other than those listed, the user must provide accurate conversion data for safe operation.
The Convectron gauge is not intended for use above 1000 Torr true pressure.