background image

© Bronkhorst High-Tech B.V.

5

7.2

BUS/digital operation

For this procedure see description 
for RS232 operation or specific 
(optional) fieldbus. 
RS232 connection cable 7.03.444 
enables to use (free downloadable) 
tooling programs for Windows.

Example 3 

Example 4 

8

Multi-functional switch operation

Using the 2 colored LEDs and the switch on the 

mini CORI-FLOW™

, several

actions can be monitored and started. The green LED is used for status
indication. The red LED is used for errors/warnings. The switch can be used to
start several actions, such as auto-zero, restore factory settings and bus-
initialization actions. See specific zero-procedure part in this manual or see
‘General instructions for 

mini CORI-FLOW™

’ for more details.

Important warning!

 When red LED blinks shortly: measuring signal is unstably or noisy: instrument is possibly

exposed to vibrations or pulsating flow.

9

Purging

In systems for use with corrosive or reactive fluids, purging with an inert gas is absolutely necessary before use. After
use with corrosive or reactive fluids, complete purging is also required before exposing the system to air. Purge the 

mini CORI-FLOW™

 with actual fluid to get rid of all the Air/gas in the tubes (for liquids) or to remove all possible

condensation drops (for gases).

Special control mode for purging

In case of a 

 mini CORI-FLOW™

 controller give setpoint = 100 % to control the valve or the pump. It is also possible

to use special control mode = 8 to fully open the valve or set the pump at max. rpm, using a digital interface. This
will bypass the PID-controller and might be usefull when having the 

mini CORI-FLOW™

 set to a low capacity. It will

ensure you to get the highest possible flow for purging.

Warm-up time

Let the 

mini CORI-FLOW™

 warm-up for at least 30 minutes for best accuracy.

10

Zeroing

Before first use, when process conditions change significantly (especially temperature) or when the instrument has
been re-mounted (e.g. after servicing) it is recommended to perform an automatic zero action with the 

mini CORI-

FLOW™

. This action can be started manually (as described below) or via a digital interface (see document Operation

instructions digital instruments: 9.17.023). Under normal (constant) conditions it will not be necessary to zero before
each application start-up.

10.1

Set process conditions

After warming-up, pressuring and purging the system, including the 

mini CORI-FLOW™

,

prepare the 

mini CORI-FLOW™

 for actual process conditions.

10.2

Stop flow

Make sure there is no flow through the 

mini CORI-

FLOW™

 by closing a shut-off valve (in front of and)

after the instrument. At least one (shut-off) valve
after the 

mini CORI-FLOW™

 is required. High

quality up- and downstream shut-off valves are
recommended for proper zero point calibration.

10.3

Start Auto-Zero

With no flow, use the push-button switch (#) on the 

mini CORI-FLOW™

 to start the zero adjustment

procedure (LEDs first will go off). (1), (2) and (3):
Press the push-button (#) and hold it for 8..12
seconds (each 4 seconds the LED-indication will
change). (4): Release the push-button (#) when the
green LED is on.

10.4

Zeroing procedure

The zeroing procedure will start now and the green
LED will blink fast. The procedure will take at least
45 sec. When the signal is not stable the procedure
will take longer (max. 6 retries will be performed).
The red LED will blink when signal is unstable/noisy.
Make sure that there is no flow through the 

mini

CORI-FLOW™

 when performing the zeroing

procedure and avoid vibrations !

45 sec.

10.5

Ready

When indication is showing 0% signal and the
green indication LED is continuously on, then zero
has been performed well. Instrument is in normal
operation mode now and ready for use.

Measurement check

For a controller: send a setpoint to the 

mini CORI-FLOW™

 and check the measured value.

Make sure the 

mini CORI-FLOW™

 indicates 0% at zero flow.

Your 

mini CORI-FLOW™

 is now ready for operation.

Bronkhorst High-Tech B.V., Nijverheidsstraat 1a, NL-7261AK Ruurlo  The Netherlands

T

 +31(0)573 45 88 00   

F

 +31(0)573 45 88 08   

I

 www.bronkhorst.com   

E

 [email protected]

Содержание mini CORI-FLOW

Страница 1: ...Quick Installation Guide Doc nr 9 17 052H Date 10 12 2015 ...

Страница 2: ...Plot Manual document nr 9 17 030 Depending on optional fieldbus interface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 044 FLOW BUS Interface 9 17 024 9 16 048 PROFIBUS DP Interface 9 17 025 9 16 049 DeviceNet Interface 9 17 026 9 16 050 Modbus interface 9 17 035 9 16 066 i www These documents can be downloaded from the website https www...

Страница 3: ... water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings Mount the mini CORI FLOW instrument with screws in the body to a rigid stiff base body or heavy mass such as a wall heavy rig or stable steel construction...

Страница 4: ...leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Gas tight is not the same as liquid tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using pumps which might lead to errors in measurement 6 Electrical connection Electrical connections must be made...

Страница 5: ...Special control mode for purging In case of a mini CORI FLOW controller give setpoint 100 to control the valve or the pump It is also possible to use special control mode 8 to fully open the valve or set the pump at max rpm using a digital interface This will bypass the PID controller and might be usefull when having the mini CORI FLOW set to a low capacity It will ensure you to get the highest po...

Страница 6: ...The red LED will blink when signal is unstable noisy Make sure that there is no flow through the mini CORI FLOW when performing the zeroing procedure and avoid vibrations 45 sec 10 5 Ready When indication is showing 0 signal and the green indication LED is continuously on then zero has been performed well Instrument is in normal operation mode now and ready for use Measurement check For a controll...

Отзывы: