SCOPE OF THIS GUIDE

mini CORI-FLOW™ 

instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

gases, independent of fluid properties. These smart Coriolis instruments offer multiple process values as input- or output
parameters. Many parameters can be read and/or changed using analog or digital interfaces.
Output parameters are: mass flow, density, temperature, totalized mass flow, alarms.
Input parameters are: setpoint (desired mass flow rate for controllers), reset alarm/counter.

This manual will help you start-up your 

mini CORI-FLOW™

 in only 10 steps and contains:

1. 

Instrument functional properties

2. 

Check safety properties

3. 

Check piping

4. 

Mount/install instrument

5. 

Leak check

6. 

Electrical connection

7. 

Operation

8. 

Multi-functional switch

9. 

Purging

10. 

Zeroing

mini CORI-FLOW™

 instruments have modular instruction manuals consisting of:

- Instruction manual 

mini CORI-FLOW™ 

(document nr. 9.17.050)

Info about e.g. : sensors, valves, liquid dosing systems, maintenance, tooling, calibration, Kv-value calculation, troubleshooting.

- Operation instructions digital instruments 

(document nr. 9.17.023)

- Hookup diagram 

mini CORI-FLOW™

 and 

CORI-FLOW™

 (general)

(document nr. 9.16.044)

- FlowPlot Manual

(document nr. 9.17.030)

Depending on optional fieldbus interface:

Fieldbus/interface description

Manual

Hookup diagram

- RS232 Interface with FLOW-BUS protocol

9.17.027

9.16.044

- FLOW-BUS Interface

9.17.024

9.16.048

- PROFIBUS DP Interface

9.17.025

9.16.049

- DeviceNet™ Interface

9.17.026

9.16.050

- Modbus interface

9.17.035

9.16.066

i

www

These documents can be downloaded from the website: 

http

s

://www.bronkhorst.com/qrcoriolis

 

 

or can be sent by e-mail on request.

Notes for temperature considerations

After having used the 

mini CORI-FLOW™

 the first time at low temperature tighten the fluid adaptor screws again in order to

prevent any leakage! Please note: if you do not tighten, the leaking adaptor / fitting can cause damage!
After the first shrinking and tightening of the screws, no further precaution is necessary.

Note that the maximum temperature in the housing of the 

mini CORI-FLOW™

 is 70 °C. 

To check this, the internal temperature sensor can be used. Temperature can be readout digitally via FLOW-DDE/
E-8000 (FlowDDE par. 142) or BRIGHT (local readout/control module).
Make sure the temperature value readout here (=actual temperature in housing) will not exceed 70 °C.

STARTING-UP

1

Check mini CORI-FLOW™ functional properties

Before installing your Mass Flow Meter/Controller it is important to 
read the attached label and check:

Flow rate

Fluid to be measured

Up- and downstream pressures

Input/output signal

Temperature

Valve type 

(if controller)

© Bronkhorst High-Tech B.V.

3

2

Check test-pressure

The tested pressure is stated on the instrument with a red-coloured sticker.
Before installation, make sure that the test pressure is in accordance with
normal safety factors for your application. If there is no Pressure Testing
Sticker on the device or if the test pressure is incorrect, the instrument
should

not

be mounted in the process line and be returned to the factory.

3

Check if system piping is clean

For reliable measurement always make sure the fluid stream is clean.
Use filters to assure a particle free liquid stream or a moisture- and
oil-free gas stream. Recommended pore size: 10...40 µm. If back flow can
occur, a downstream filter and a check valve are recommended too.
For high flow rates select a suitable filter size, to avoid too high pressure
drop or cavitation.

Warning!

During the manufacturing process, the instruments have been tested with water. Despite the fact that the
instruments have been purged thoroughly afterwards, we cannot guarantee that the delivered instruments are
absolutely free from water droplets. Bronkhorst strongly recommends performing an additional, adequate drying
procedure for those applications where remaining water particles may cause undesired reactions such as corrosion.

4.1

Mount/install instrument properly

Install the

mini CORI-FLOW™

Meter/Controller in the line and tighten the fittings according to the instructions of

the supplier of the fittings. Mount the

mini CORI-FLOW™

instrument, with screws in the body, to a rigid, stiff base

body or heavy mass, such as a wall, heavy rig or stable steel construction. This is essential to achieve optimal
accuracy with the

mini CORI-FLOW™

instrument.

4.2

Flow direction

Install the

mini CORI-FLOW™

in accordance with the direction of

the FLOW arrow. The arrow for flow direction is indicated on the

mini CORI-FLOW™

, between process fittings.

4.3

Base mounting

Mount the

mini CORI-FLOW™

instrument, with screws in the body, to a rigid, stiff base body or heavy mass on a

vibration-free position, such as a wall, heavy rig or stable construction. This is essential to achieve optimal accuracy
with the

mini CORI-FLOW™

instrument.

By default the

mini CORI-FLOW™ M12, M13

and

M14

will be delivered on a special mounting block for achieving

optimal accuracy.This mounting block has a mass and stiffness precisely tuned for the specific model.

Removing the mounting block will cause inaccuracy unless the instrument is firmly mounted on a stiff and rugged
surface.

4.4

Mounting position general

For gas and liquid

mini CORI-FLOW™

meters can be mounted in any position for a proper measurement.

Содержание mini CORI-FLOW

Страница 1: ...Quick Installation Guide Doc nr 9 17 052H Date 10 12 2015 ...

Страница 2: ...Plot Manual document nr 9 17 030 Depending on optional fieldbus interface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 044 FLOW BUS Interface 9 17 024 9 16 048 PROFIBUS DP Interface 9 17 025 9 16 049 DeviceNet Interface 9 17 026 9 16 050 Modbus interface 9 17 035 9 16 066 i www These documents can be downloaded from the website https www...

Страница 3: ... water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings Mount the mini CORI FLOW instrument with screws in the body to a rigid stiff base body or heavy mass such as a wall heavy rig or stable steel construction...

Страница 4: ...leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Gas tight is not the same as liquid tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using pumps which might lead to errors in measurement 6 Electrical connection Electrical connections must be made...

Страница 5: ...Special control mode for purging In case of a mini CORI FLOW controller give setpoint 100 to control the valve or the pump It is also possible to use special control mode 8 to fully open the valve or set the pump at max rpm using a digital interface This will bypass the PID controller and might be usefull when having the mini CORI FLOW set to a low capacity It will ensure you to get the highest po...

Страница 6: ...The red LED will blink when signal is unstable noisy Make sure that there is no flow through the mini CORI FLOW when performing the zeroing procedure and avoid vibrations 45 sec 10 5 Ready When indication is showing 0 signal and the green indication LED is continuously on then zero has been performed well Instrument is in normal operation mode now and ready for use Measurement check For a controll...

Отзывы: