12. Enter the desired deviation alarm tolerance. The deviation alarm is an
alarm point that changes with the temperature set point. If this alarm point
is exceeded, the alarm horn sounds and ALARM displays in the zone
operating screen.
13. Enter the desired shutdown alarm tolerance. The shutdown alarm is an
alarm point that changes with the temperature set point. If this alarm point
is exceeded, the alarm horn sounds and the controller aborts the cure cycle.
A notification of the cycle abort displays in the zone screen.
14. Enter the vacuum high and low alarm set points. The alarm sounds and
ALARM displays in the zone operating screen if either of these points are
exceeded. Exceeding either of the vacuum alarm points does not cause the
controller to abort the cure cycle.
15. Specify active and inactive thermocouples. Active thermocouples are
indicated by a check mark in the check box next the corresponding
thermocouple number. Touch the check boxes to remove the check mark
from the inactive thermocouples.
16. Press the CONTINUE TO STEP 3 button.
© BriskHeat
®
Corporation. All rights reserved.
18
ACR
®
3 MiniPRO™ Hot Bonder
18. Fill in all fields to the desired setting. Touch FINISH when complete.:
Ramp Rate:
the heat up rate in degrees per minute from starting
temperature to Dwell Temperature.
Dwell Temp:
the temperature the program is to hold for the
programmed length of time
Dwell Time:
the length of time to hold at the programmed Dwell
Temperature.
Cool Down Rate:
the rate in degrees per minute from the last dwell
temperature to the End of Cycle Temp.
NOTE: Since the Hot Bonder cannot force an object to cool down faster than
the physical conditions will allow, both high temperature alarms are
disabled during the cool down phase.
End of Cycle Temp:
the temperature at which the cure is considered
complete.
NOTE: The end of cycle temperature must be above the ambient
temperature for it to complete.