4 DISPLAY
(FIG 2)
(FIG 3)
(FIG 4)
(FIG 5)
T*32.1C
H 35% L/
High Temperature Alarm Display (FIG 2)
– If the measured temperature inside the incubator goes up by more than the
figure in the ALARM HI screen, the alarm will sound immediately and “H” will be displayed. Press OK to silence the alarm for 30
minutes.
If the high temperature problem rectifies itself the “H” remains on the display to show this has happened. Press OK to clear
the indicator. Check the incubator is not (and has not been) in direct sunlight or too near a heat source such as a room heater.
It is advisable to candle eggs a number of times after this event to check for losses.
Low Temperature Alarm Display (FIG 3)
– If the measured temperature inside the incubator goes down by more than
the figure in the ALARM LO screen, after 60 minutes “L” will be displayed and the alarm will sound. Press OK to silence the alarm for
30 minutes.
If the low temperature problem rectifies itself the “L” remains on the display to show this has happened. Press OK to clear the
indicator. Check the incubator is not (and has not been) in a cold draught and that the room temperature has not dropped
significantly. It is advisable to candle eggs a number of times after this event to check for losses.
Low Room Temperature Alarm (FIG 4)
– If the calculated room temperature remains too low for optimum results for
more than 1 hour a warning is displayed “-RM” and an alarm will sound. Press OK to silence the alarm for 30 minutes.
If the low temperature problem rectifies itself the “-RM” remains on the display to show this has happened. Press OK to clear
the indicator.
Check the incubator is not (and has not been) in a cold draught and that the room temperature has not dropped significantly.
It is advisable to candle eggs a number of times after this event to check for losses.
High Room Temperature Alarm (FIG 5)
– If the calculated room temperature remains too high for optimum results for
more than 1 hour a warning is displayed “+RM” and an alarm will sound. Press OK to silence the alarm for 30 minutes.
If the high temperature problem rectifies itself the “+RM” remains on the display to show this has happened. Press OK to clear
the indicator.
Check the incubator is not (and has not been) in direct sunlight or too near a heat source such as a room heater. The eggs
themselves create significant metabolic heating at later stages of incubation and may contribute to this if room temperature is
high. It is advisable to candle eggs a number of times after this event to check for losses.
T*37.5C
-RM /
T*37.5C
+RM /
T*39.8C
H 35% H/