background image

GP100 and GP250 User Instructions

13

   

2.6 Scale Communications

1.

Turn the scale on.

2.

Press and hold the 

[TARE]

 key for 3-5 seconds. The scale will display 

TARE

SET

,

 VERSION NUMBER

 and 

UNIT

.

3.

Release the 

[TARE]

 key when you see 

UNITS

 on the display

4.

Press the 

[UNITS]

 key until 

Port

 on the display.

5.

Press the 

[TARE]

 key and then press the 

[UNITS]

 key to toggle to the desired 

software protocol.

6.

Once the correct selection is displayed, press the 

[TARE]

 key to accept.

7.

With 

Port

 on the display, press the 

[UNITS]

 key. 

Baud rate

 will be displayed.

8.

Press the 

[TARE]

 key and press the 

[UNITS]

 key to toggle to the desired baud 

rate communication setting needed for your software.

9.

Press the 

[TARE]

 key and press the 

[UNITS]

 key. 

Par

 will be displayed.

10.

Press the 

[TARE]

 key and press the 

[UNITS]

 key to toggle to desired parity 

format.

11.

Press the 

[TARE]

 key to accept the selection. 

Par

 is displayed.

12.

Press the 

[UNITS]

 key and 

End 

will be displayed.

13.

Press the 

[TARE]

 key to exit the menu structure.

NOTE: For UPS Worldship software, you will need to select the NCI protocol, 
4800 baud rate, 7 even parity in the scale menu configuration. In the actual UPS 
software, select the 3835 for type of scale.

Содержание GP100-USB

Страница 1: ...GP100 USB and GP250 USB Electronic General Purpose Bench Scale User Instructions AWT 35 500825 Issue AC...

Страница 2: ...wo dimensional work and the making of a copy in two dimensions of a three dimensional work stored in any medium by electronic means or transmitted in any form or by any means including electronic mech...

Страница 3: ...atures 5 Installation 6 Battery Installation 6 Routine Maintenance 7 Warranty 7 Cleaning and Care 7 Chapter 2 General Operation 8 Key Functions 8 Zero ON OFF 8 Tare 8 Units 9 Hold 9 Auto Power Off Con...

Страница 4: ...of the pages 1 1 1 Text Conventions The keys used to interface with the GP100 250 are located on the front panel of the indicator The keystrokes are shown in BOLD between brackets e g ZERO Displayed m...

Страница 5: ...the scale while using an empty container to hold them in place on the scale Simply place the empty container on the scale press the TARE key the scale will now display 0 0 weight Place items in the em...

Страница 6: ...tly filed with the shipper 1 Open the box and carefully remove the scale from the shipping carton save the shipping carton for future use 2 Place the scale on a flat level surface free from vibration...

Страница 7: ...and Care l Handle the scale with care Dropping the scale may cause damage to the scale which will not allow the scale to weigh properly This type of damage is considered abuse and violates the product...

Страница 8: ...weight or a negative weight appears on the display 2 1 2 Tare To use the tare feature 1 Place an empty container on the weighing platform 2 Press the TARE key The scale will display zero weight When a...

Страница 9: ...active 3 Place an item on the scale wait two seconds for the weight to become stable 4 When the weight becomes stable the Hold annunciator will no longer flash 5 Remove the item from the scale The we...

Страница 10: ...times The display will read END 7 Press the TARE key to exit the setting and return the scale to 0 0 weight 2 4 Scale Calibration Periodically your scale may require calibration to ensure proper weigh...

Страница 11: ...ble 2 1 Calibration Weight Table 2 2 Span Range 2 5 Connection with a PC The scale can automatically connect to a PC via a USB cable Note The USB cable is not included with the scale A USB A to USB B...

Страница 12: ...nstall Update the USB Driver Use the link below to access the USB Installation Guide The document provides detailed installation and basic operation procedure and setup http www brecknellscales com do...

Страница 13: ...press the UNITS key Baud rate will be displayed 8 Press the TARE key and press the UNITS key to toggle to the desired baud rate communication setting needed for your software 9 Press the TARE key and...

Страница 14: ...has occurred Calibrate the scale see page 10 for instructions After pressing the HOLD key and placing a weight on the scale the weight will not hold on the display The continues to flash Remove weigh...

Страница 15: ......

Страница 16: ...38 8702 Fax 507 238 8271 Email sales brecknellscales com http www brecknellscales com Brecknell UK Foundry Lane Smethwick West Midlands England B66 2LP Tel 44 0 8452 46 6717 Fax 44 0 8452 46 6718 Emai...

Отзывы: