9 Maintenance Instructions
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
Inspection Equipment and Tools
The following equipment is required to perform an inspection on the
Clearway 2:
•
Gas flow analyser.
or
•
Manometer capable of measuring at least ±70 cmH
2
O (± 68.6 hPa) (±
0.75%) and a flow meter capable of measuring at least 300 sl/m (± 2%).
•
70% Alcohol cleaning wipes
•
Test Lung (part no 6011)
•
6mm leak port (part no 00997)
•
Clearway Breathing Circuit (part no 0960.101)
•
Medical Electrical Safety tester
The following tools are required to open the case, remove and replace
components when necessary:
•
Torx screwdriver T15 x 200mm (To open the case)
•
Pozidrive screwdriver No. 1
•
Hex screwdriver 2mm
•
Hex screwdriver 2.5mm (To open the battery cover)
•
Soldering iron, to remove and replace the alarm battery every 5 years
Содержание Clearway 2
Страница 1: ...Service Manual Technical Instructions for Use 2210 2657 Revision 3 0...
Страница 25: ...25 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB...
Страница 26: ...Part identification 26 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Inside Rear Case...
Страница 27: ...27 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Power PCB...
Страница 30: ...Functional Diagrams 30 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Control PCB block diagram...
Страница 31: ...31 Functional Diagrams NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB block diagram...
Страница 32: ...Functional Diagrams 32 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Pneumatic diagrams...