13 Maintenance Instructions
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
Remove the test lung and replace with a 6mm leak (part no. 00997).
Use the front switch to activate Exsufflation (-) a number of times
Compare the value on the flow meter with the CPF value displayed on screen
(value updates when the switch is released.)
Displayed and measured values must be within ± 10 % of each other.
Hand control (when used)
Attach the breathing circuit (part 2210.101) and test lung (part 6011) to the
patient outlet.
Select Manual Mode and ensure that the remote hand control switch activates
insufflation and exsufflation appropriately. The Start/Stop button must start
or stop a treatment in the Auto Modes only.