21 Service Menu
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
7.
Press Next to complete.
Valve Calibration
The following procedure is factory set and not required unless a valve assembly
or either PCB has been replaced or a software upgrade has been performed.
1.
Enter the service menu and select ‘valve calibration’.
2.
Do not connect anything to the patient outlet.
Press ‘ZERO’ then ‘Next’.