Manual C118 EN Rev04 

04.2017 

12 

No reliable measurement can be expected unless the input 

signal is clearly accepted!

Note:

The input indicator may still be operational with the power 

supply no longer adequate.

The 

indispensable 

pre-condition

to reliable 

measurements

MAX/MIN/AVG Function 

The C118 detects and memorizes the highest (MAX) and 

the lowest (MIN) measurement within a controlled period 

of time. From those, it further computes the average (AVG) 

of all measurements.  

Begin and end of this evaluation period is controlled, under 

continued measurement, by the BEGIN/END key. First ac-

tuation start memorizing, next one stops it. Display 

FUNC/RUN indicates the function as being active. Once it 

has been terminated (power left on), the results can be re-

called to display by the key MAX/MIN/AVG. Power OFF 

switching cancels the memorized values. 

Note: Do not change the measuring mode or other terms 

during the evaluating period.

MAX/MIN/AVG 

memorizing

Time Base of Readings  

The facility of the C118 converts the originally measured 

signal frequency into readings with a corresponding time 

base, as indicated at the display. Select desired time base by 

the key marked /h /min /sec /SF. The conversion has no in-

fluence to the measuring sequence, nor to the updating se-

quence of the display.

Totalizing 

SF evokes a totalizing function within the instrument. It is 

started and stopped by the BEGIN/END key, as described 

above. Starting simultaneously cancels the previous result. 

In FIX mode, the display reads the incoming signal pulses, 

reduced by the input pre-divider as programmed by P09. 

In VAR mode, they are further scaled by the computing 

factor, as programmed in steps No. 02 … 04. The pro-

grammed mode, multiplying or dividing, and the factor, re-

sult in correspondingly enlarged or reduced reading steps 

per input pulse. If the corresponding reading step thus re-

sults in a fraction, a decimal point in display can be set by 

program step P10.  

The totalizer capacity is up to 99999. Overflow will cause 

errors.  

The analog output cannot be used in this SF mode.

Time base of 

speed readings

/h 

 /min

/sec

 SF = totalizing 

function

Manual C118 EN Rev04 

04.2017 

13 

Operating Modes FIX and VAR

This is one of the facilities making the C118 that much  

useful. Select between both modes by actuating the 

FIX/VAR key. VAR mode is indicated by a sign in the  

display. Programmed parameters will not be deleted when 

switching to FIX.

FIX = straight 

tachometer

VAR = using the 

computing 

facilities 

FIX mode 

Measuring the rotational speed, with these firm characteris-

tics: 

Calibrated to 1 pulse per revolution. Minimum measuring 

time = 1 sec. Low end = 30 pulses / min. 

Readings in terms of /sec, or /min (RPM), or /hour, as se-

lected by this key.  

FIX mode uses a floating decimal point: 

Starting with one decimal at low speed. With measurements 

above 10000.0, the last digit is kept on zero, to avoid flick-

ering readings. The decimal point is cancelled when the 

measurement exceeds 99999.0.  

Note: The thus achieved decimal selection will be main-

tained with decreasing measurements. Return to the initial 

decimals sequence by switching the instrument off/on. 

Totalizing (SF) may be used, counting the input pulses, not 

scaled, but reduced by the preset input pre-divider. 

Further effective the MAX/MIN/AVG memory. Also the 

analog output, using the characteristics, as programmed.

FIX

characteristics

VAR mode 

This mode activates all the capabilities included into the 

MOVIPORT C118, by programmable parameters, as de-

scribed in the subsequent chapters. 

Programming can be done at any time, and in any mode. 

The thus defined functions become effective in the VAR 

mode only.  

Note: 

This feature assigns two readings to one measurement, se-

lected just by switching between FIX and VAR. For in-

stance, the rotational speed by FIX, and the circumferential 

speed by VAR as computed from its circumference. Or, the 

measured input speed of a gear in FIX, and the output speed 

in VAR, as computed from its turn down ratio. 

VAR 

mode facilities

Programming 

procedure and  

parameters list 

on page 14

Two readings 

utilizing one 

measurement  

Содержание MOVIPORT C118 Series

Страница 1: ...Universal Hand Held Tachometer MOVIPORT C118 Series Manual Hand Held Tachometer C118 Cable SAK 2m Sensor A1S30P95...

Страница 2: ......

Страница 3: ...put Response 11 MAX MIN Memory 12 Totalizing Function 12 FIX VAR Operation Modes 13 Programming Procedure and List of Steps 14 Parameter Functions Shortest and Maximum Measuring Time Allowance 15 Comp...

Страница 4: ...6P95 with 5m firmly attached cable in place of A1S30P95 SAK 2m Equipment otherequipment as under C118 2BS C118 3BP Instrument with RS232 Data Interface and adapter cable L3D01 otherequipment as under...

Страница 5: ...able position VAR mode Memory Maximum Minimum Average over a controllable period of time Call to display by push button Signal Input To pulses or AC voltage from 50 mVeff up to 60 volts Selection of a...

Страница 6: ...sitivity C118 2 and C118 3 only Jack to supply unit U1H008 Display reading measurements with their time reference operating mode pro gram Step No and its parameter Key board to select the functions se...

Страница 7: ...gramming procedure and function in structions start on page 14 plug in sensor immediately into socket or use connection cable For normal operation use predetermined automatic mode Manual knob preset a...

Страница 8: ...ape U1A006 A1S37P Sensing wheel twin type wheel 63 mm 1000 pulses meter for length detection Speed measuring range 0 1200 m min A1L04B200 Sensing wheel twin type wheel 159 mm 1000 pulses meter for len...

Страница 9: ...faster reaction at a slow motion they should be placed with equal distances be tween them The acceptable scanning distance clearance widely de pends on the contrast of the mark In general it will be 0...

Страница 10: ...range from 40 C 125 C Magnetic induction type sensors are active but limited to rather high speed as in turbo blower engines Further to detect AC magnetic fields as in the neighborhood of igni tion co...

Страница 11: ...ccepts the other sensors see page 6 too Please observe the guiding slot at the socket The screw at the connector will be downside For other signal sources the ready made cable L2A12PS 2m may be used o...

Страница 12: ...wered by batteries cells or from mains by supply unit Battery replacement Power ON OFF The ON OFF key switches power to the instrument Further it may be programmed to automatically switch off by itsel...

Страница 13: ...adjustment of the maximum sensitivity Model Nos C118 2 and C118 3 only The fast automatic signal conditioning might lose its cor recting function when the signal frequency drops below ap proximately 1...

Страница 14: ...Base of Readings The facility of the C118 converts the originally measured signal frequency into readings with a corresponding time base as indicated at the display Select desired time base by the key...

Страница 15: ...ill be main tained with decreasing measurements Return to the initial decimals sequence by switching the instrument off on Totalizing SF may be used counting the input pulses not scaled but reduced by...

Страница 16: ...in both modes FIX and VAR Programmed characteristics however get valid in VAR mode only In FIX mode all functions are as listed on page 5 Press key PROG to enter the programming phase Step 00 appears...

Страница 17: ...ly by keys or Step No 00 minimum measuring time allowance determines tracking sequence but also steadiness of readings Maximum time allowance for one measurement This program step sets a limit to the...

Страница 18: ...the di viding factor Set parameter of step 02 to 1 to achieve the dividing mode Determining the relation between measurement and reading Step No 02 Mode of computing factor 0 multiplying 1 dividing D...

Страница 19: ...tput High End 4 volts Assignment of the Analog Output The parameter defines the displayed value assigned to the high end 4 volts of the output however it may be achieved disregarding the s min h refer...

Страница 20: ...tion will divide the incoming pulse rate accordingly This returns the pulse se quence to one pulse per revolution what obviously ensures the most accurate repetition as the best basis for stable measu...

Страница 21: ...Minimum Time Allowance 0 50 s 01 Maximum Time Allowance 8 00 s 02 Computing Factor Mode 0 multiplying 03 decimals none 04 numerical 000001 05 Display decimals none 06 Analog Output high end 010000 07...

Страница 22: ...d results or sent only once after request Transmission Request message syntax STX n H z ETX The type of measurement required is defined by a figure n in the syntax n 1 actual most recent measurement 2...

Страница 23: ...de 02 ETX end of text in HEX code 03 CR carriage return in HEX code 0D Connection cable L3D01 and Data Port Connector type of cable L3D01 USB Mini B male to 9 pole SUB D male pin No signal 2 RxD m 3 T...

Страница 24: ...Manual C118 EN Rev04 04 2017 22 4 2017 D 71334 Waiblingen Esslinger Str 26 Phone 49 0 7151 956230 Fax 49 0 7151 956250 E Mail info braun tacho de Internet www braun tacho de...

Отзывы: