System User Guide Rev. 1.1, NaviVision (Hardware Package to Upgrade Siemens C-Arm) Ver. 1.0
77
NaviVision System
The use of accessories or cables other than those listed above may result in in-
creased emissions or decreased immunity of the
NaviVision
system.
where
P
is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and
d
is the recommended
separation distance in metres
(m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
should be less than the compli-
ance level
b
in each frequency
range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can-
not be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the upgraded
system is used exceeds the applicable RF compli-
ance level above, the upgraded system should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating
the upgraded
system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.