6
BOWA SHE SHA smoke evacuation unit
MN009-940-S0 EN
10. This equipment needs special precautions regarding Electro Magnetic Compatibility and needs
to be installed according to EMC information found in this manual.
11. This equipment utilizes mobile RF communications equipment that can affect medical electrical
equipment.
12. This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which
case the user will be required to correct the interference at their expense.
13. This equipment operates in the following radio frequency specifications:
RX modulation: Pulse-width coded, AM 100% modulation
TX Frequencies: Manchester encoded,
A = fc = / -423.75 kHz, B = fc +/- 484.29 kHz
Low bit: transition A to B
High bit: transition B to A
14. To isolate equipment from supply mains, unplug the power cord from the appliance inlet on the
unit or receptacle. Position the equipment to allow for ease of unplugging power cord.
15. Potential Equalization Conductor: Terminal located on back panel for connection of potential
equalization. Conductor complies with requirements per IEC 60601-1 (2005).
The SHE SHA Smoke Evacuation System(s) and all filters are not intended for contact with patients.