BOWA SHE SHA smoke evacuation unit  

MN009-940-S0 EN 

 

Reading the Filter Life Indicator: 
Install an unused SHE SHA filter into the system per the installation instructions contained in this 
operator’s manual. When the system is turned on, the filter life indicator will light up the top orange 
filter life indicator LED, indicating 100% filter life. The indicator will progress through vertically 
allocated LED and the lowest LED will begin flashing RED to indicate that the filter has expended its 
useful life and requires replacement. 

 

When the maximum filter life is consumed and the smoke evacuator is not powered off for greater 
than six (6) hours or the main power is disconnected – a new filter is required to activate the smoke 
evacuator and make operational. 

 

• 

FUSES 

(circuit board)

 

Two 8 AMP fuses are located on the circuit board within the housing of the system. It electrically 
protects both the system and the operator from damage or injury. If the system is overheated or if 
there is an electrical surge in the electrical system, fuses will break and the system will not operate.   

 

When the Service light illuminates, please contact BOWA’s technical service for system service 
instructions, see chapter 4 technical service. 

 

•  OCCLUSION

 DETECTION

 

 

 
The SHE SHA unit provides an occlusion detection feature to avoid undesired tissue suction for 
maximum patient safety. The LED codes shown in the table below indicate if the occlusion detection 
is enabled or disabled and if an occlusion is detected. 

 

LED Functions 

Meaning and corrective measures 

On LED 

Standby LED 

 

 

 

Unit in Standby, Occlusion Detection disabled 

 
 

 

 

Unit in Standby, Occlusion Detection enabled 

 

 

Unit running, Occlusion Detection disabled 

 

 

 

 

Unit running, Occlusion Detection enabled 

 

 

Unit is stopped due to a detected Occlusion 

  Must be restarted by pressing On / Standby button or 

remote switch activator 

 

 

 

Содержание SHE SHA

Страница 1: ...SHE SHA SMOKE EVACUATION UNIT OPERATING MANUAL...

Страница 2: ...OCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 ANSI AAMI ES60601 1 2005 3rd ed CAN CSA C22 2 NO 601 1 AND CAN CSA C22 2 No 60601 1 2008 9D93 This device complies with Part 15 of th...

Страница 3: ...utions 8 2 Operating Instructions 10 2 1 User Interface Components 10 2 2 SHE SHA Filter Instructions 13 2 3 Set up and Operation 15 2 4 Technical data SHE SHA smoke evacuation system 16 2 5 Electroma...

Страница 4: ...The first stage filtration is a prefilter whose function is to trap and remove gross particulate and casual fluid The second stage filtration is ULPA grade Ultra Low Penetration Air filter whose high...

Страница 5: ...th domestically and internationally 1 The BOWA SHE SHA Smoke Evacuation System s complies with IEC60601 1 electrical specifications in the following systems 220 240 VAC 50 60 Hz 2 Type of protection a...

Страница 6: ...uction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required t...

Страница 7: ...ion see chapter 2 1 occlusion detection To maximize patient safety the tubing or wand should not come into direct contact with tissue Otherwise patient injury may result The BOWA SHE SHA filters and s...

Страница 8: ...ssembly Failure to install a fluid collection device could cause filter blockage and electrical damage The SHE SHA filter should be changed according to the life of the filter The SHE SHA filter used...

Страница 9: ...do not remove cover refer servicing to qualified service personnel Caution Type CF applied part IXP1 Protection against ingress of water as detailed in IEC 60529 Alternating current Protective earth...

Страница 10: ...the suction on off switch suction power adjustment filter life indicator and service indicator light 1 ON OFF button 2 Orange Standby LED 3 Green fan ON LED 4 Speed Select Button 5 Suction Levels LED...

Страница 11: ...E SHA Smoke Evacuation System also comes equipped with a pneumatic foot switch and remote switch activator Connect the cable of the foot switch to the corresponding socket on the front panel When the...

Страница 12: ...board within the housing of the system It electrically protects both the system and the operator from damage or injury If the system is overheated or if there is an electrical surge in the electrical...

Страница 13: ...dby LED starts blinking if the occlusion detection is enabled The Orange Standby LED stops blinking if the occlusion detection is disabled 2 2 SHE SHA Filter Instructions BOWA Filter REF 951 001 Confi...

Страница 14: ...y cause filter blockage and or electrical damage Filter Removal Instructions 1 After the filter has been exhausted and requires changing turn the Smoke Evacuation System off and disconnect any accesso...

Страница 15: ...n tubing is fully seated in the inlet of the filter Route tubing in such a way as to minimize potential trip hazard by users or patients 5 Activate the power unit by a Pressing the suction ON OFF swit...

Страница 16: ...LPM 2 CFM Dimensions H x W x D 152 mm x 279 mm x 394 mm 6 x 11 x 15 5 Weight 4 4 kg 5 5 kg incl filter Noise Level dBA max 55 0 dBA Foot switch Pneumatic Standard Remote Control Activation YES Safety...

Страница 17: ...ctromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The SHE SHA uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interf...

Страница 18: ...hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environ...

Страница 19: ...bles than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Radiated RF 3 V m d 1 7 P 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz 3...

Страница 20: ...AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should...

Страница 21: ...by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the model SHE SHA as recommended below according to the maximum output power of the communica...

Страница 22: ...ast every year This inspection should include checks for Damage to the power cord Damage to the power plug or power inlet module Proper mating cleanliness and absence of damage to the filter inlet Obv...

Страница 23: ...4 Call BOWA Technical Services 2 Smoke Evacuation System does not function even though suction ON button is depressed 1 Not plugged into an electrical outlet 2 Fuses are blown 3 Electronic system fail...

Страница 24: ...r com NL 2513 BH The Hague The Netherlands Distributed by BOWA electronic GmbH Co KG Heinrich Hertz Stra e 4 10 72810 Gomaringen Germany Phone 49 0 7072 6002 0 Fax 49 0 7072 6002 33 info bowa medical...

Отзывы: