Risk of injury due to heat generated by friction!
The user must ensure that as little friction heat as possible is generated during a procedure on a patient. High speeds and
high application pressure may cause thermal necrosis of tissue.
Risk of injury!
The distal end of the cutting tube is extremely sharp.
Precautions:
• Handle the cutting tube particularly cautiously and carefully.
• Where possible, have the Morcellator with protective tube in protected position (NO CUT), particularly if the cutting
tube is not used for any period.
• Insert and remove the protective tube and cutting tube under visual control (endoscopy).
Risk of injury!
When inserting the Obturator through the percutaneous access, there is risk of injury to blood vessels, intestinal loops or
the bladder. This can result in significant complications, which will require a Laparotomy.
Risk of damage to the blade of the cutting tube!
If the jaws of the gripper are not completely closed when it is pulled from the protective tube, the blade will be damaged
(see pictures).
Risk of injury!
Damaged blades and cutting tubes must not be used.
Cutting tubes must not be sharpened!