Instructions for use, ERGO 300 - BOWA- IFU-12930- ERG0300-S0-EN-20210701
5
Before use
The ERGO 300 – Morcellator and all required accessories and instruments must be placed on a level nonslip surface. Do
not allow the operating range of the device with cable to be restricted by limiting factors. The display, keypad and indicator
lights must be fully visible at all times.
It is very important to ensure that no objects can fall on the pedal. The power plug at the rear of the device must be acces-
sible at all times.
5.1
Potential equalization connection according to DIN 42801
At the back of the device a potential equalization plug is installed, according to DIN 42801.
The additional potential equalization has the task of equalizing potentials between different parts of conduc-tive materials
that can be touched at the same time, or reducing potential differences.
This connection must be used, to protect the patient, the user and third parties from touch voltages.
The equipotential plug is marked with the following symbol:
5.2
Connection to the power supply
• Before plugging the power cable into the power socket for the first time, you must check the supply voltage setting
next to the power switch.
• In order to prevent the risk of electric shock, the device may only be connected to a power network with a
PE protective earth conductor.
If the voltage shown does not correspond to the local mains voltage, the grey fuse holder must be set to the correct voltage:
1. Switch the device off.
2. Unplug the power cable.
3. Use a screwdriver to open the fuse slot.
4. Remove the fuse holder.
5. Remove the grey fuse holder and reinsert it so that the local mains voltage setting is shown in the small window.
6. Slide the grey fuse holder back in and close the fuse slot.
7. Check the mains voltage shown on the fuse slot.
8. Plug the power cable back into the device.
5.3
Device preparation
The device is not sterile on delivery. All sterilizable parts must be sterilized before use (see Chapter 7 Instrument prepara-
tion).
If the accessories have already been sterilized: when removing them from the sterile packaging, ensure that the sterile
packaging is not damaged and that the sterility indicator confirms sterility and shelf life of the sterile item has not ex-
pired.
The Morcellator (as per
option 1
to
option 3
) must be assembled under sterile conditions (wear gloves and a mask, place
Morcellator and accessories on a sterile surface).
Содержание ERGO 300
Страница 1: ...EN ERGO 300 Morcellator Information for use 12930 S0 EN...
Страница 2: ...2 Instructions for use ERGO 300 BOWA IFU 12930 ERG0300 S0 EN 20210701...
Страница 43: ...Instructions for use ERGO 300 BOWA IFU 12930 ERG0300 S0 EN 20210701 43 Appendix...
Страница 48: ......
Страница 50: ......
Страница 51: ......