1 689 989 473 | 2021-08-24
Robert Bosch GmbH
458922-31_P
al
1
2
3
52:
Removing the protective cover
10. Remove the protective backing all around the edges
on both sides of the new protective cover. Do not re‐
move the protective backing entirely.
11. Place the protective cover in the edge protection
frame and align it. Make sure that the label with the
replacement date is visible and legible from outside.
12. Place the mountings in position.
13. Fasten the mountings with the hex nuts.
14. Remove the protective backing on the inside.
15. Place the edge protection frame with the protective
cover into the protective door.
16. Fasten the edge protection frame with the protective
cover by means of the countersunk bolts (10x).
17. Remove the protective backing on the outside.
---Separator---
5.3
Checking the measuring accuracy
The measuring accuracy of the DCI 700 must be
checked by the EP test service every 2 years. The EP
test service has the accessories needed to check the
measuring accuracy and is a requirement for certifica‐
tion to ISO 9000.
---Separator---
5.4
Spare and wearing parts
Denomination
Part number
1 x test oil filter
<)
1 687 434 083
1x inlet filter, complete
1 687 434 081
1x test oil filter for inlet filter
<)
1 687 434 084
8x O-ring 28 x 2 mm for inlet filter
<)
1 680 210 179
1x thrust ring
1 680 200 102
1x protective cover
1 680 113 054
1x hose assembly for the injected-fuel quanti‐
ty
<)
1 687 410 056
1x hose assembly for the return quantity
<)
1 680 712 382
1x return adapter for Bosch CRI
<)
1 687 016 215
1x return adapter hose assembly for Bosch
CRI 2.18 and CRI 2.20
<)
1 683 351 066
1x return adapter hose assembly for Bosch
CRI piezo
<)
1 680 712 315
12x O-ring 9.5 x 2 mm for return adapter
1 687 016 215 and return adapter hose as‐
sembly 1 683 351 066
<)
1 687 016 045
2x adapter nut M12 x 1.5
<)
1 683 345 065
Denomination
Part number
2x adapter nut M14x1.5
<)
1 683 345 066
1x adapter cable for Bosch CRI 1, CRI 2x and
CRI Piezo
<)
1 684 465 900
1x adapter cable for Bosch CRI 3-25 and
CRI 3-27
<)
1 684 465 895
1x threaded plug
<)
1 683 462 169
1x iinjection chamber device
1 685 200 172
1x knurled nut D9
<)
1 683 373 086
1x knurled nut D7
<)
1 683 373 087
1x O-ring insert D9 without O-rings
<)
1 680 109 140
1x O-ring insert D7 without O-rings
<)
1 680 109 141
6x O-ring 12 x 2 mm for O-ring insert
1 680 109 140 and 1 680 109 141
<)
1 680 210 123
6x O-ring 9 x 3 mm for O-ring insert
1 680 109 140
<)
1 680 210 132
3x O-ring 7 x 4 mm for O-ring insert
1 680 109 141
<)
1 680 210 124
1x scale plate
<)
1 688 001 074
Pressure limiting valve
<)
–
1)
Work area light
<)
–
1)
---
<)
Wearing part
1)
Customer service performs maintenance and exchanges.
---
5.5
Faults
---Separator---
5.5.1
Problems performing functions
Fault
Cause and remedy
The message
Disconnected
is displayed
by the "DDM" software.
The message "Disconnected" is displayed in the "DDM" software
After the DCI 700 is switched on, the
screen remains dark and the operating
display does not light up on the monitor.
Touchscreen not functioning.
After the injector has been removed, a
large amount of test oil emerges from
the injection chamber.
After the injector has been removed, a large amount of test oil emerges from the injection chamber
Test oil emerges at the knurled nut on
the injection chamber during the test.
Test oil emerges at the knurled nut on the injection chamber during the test
The values for the amount of test oil are
too high when injectors are tested.
The values for the amount of test oil are too high when injectors are tested
The values for the amount of test oil are
too low when injectors are tested.
The values for the amount of test oil are too low when injectors are tested
No amounts of test oil are measured
when injectors are tested.
No amounts of test oil are measured when injectors are tested
---Separator---
The message
Disconnected
is displayed in "DDM."
The Internet connection has been interrupted, and the
"DDM" (Diagnostics Download Manager) software cannot
establish a connection to the DDM server.
1. Check if the network cable is intact and connected
properly.
2. For additional information, see "DDM Help Center."
3.
and install the latest DDM version.
---
en | 24 | DCI 700