background image

D7050DH 

|

 

Installation Instructions | 8.0 

Maintenance 

 

Bosch Security Systems, Inc. | 9/06 | 49325E 

 

7.3 Magnet 

Test 

 

 

To perform the Magnet Test, the control 
panel must be in the Fire Walk Test Mode.  
If the control panel is not in this mode, it 
sends fire reports to the central station.  
Refer to your control panel’s reference 
guide for Fire Walk Test information. 

1.

 

Hold a magnet horizontally against the recess in 
the D343 cover for approximately 10 sec.  

2.

 

Observe the LED. If the D7050DH is within the 
factory marked calibration range, it goes into alarm 
and the Alarm LED flashes at least once per 
second. 
a.

 

If the D7050DH is too sensitive, the Alarm 
LED flashes six times rapidly (once every 0.5 
sec) and the detector goes into alarm. 

b.

 

If the D7050DH is not sensitive enough, the 
Alarm LED flashes four times slowly (once 
every 2 sec) and the detector goes into alarm. 

3.

 

If the D7050DH is not operating, it does not signal 
an alarm. Return the detector for repair. 

7.4 

Voltage Measurement Test 

1.

 

Plug a D1005 Test Cable (optional) into the 
calibration voltage pin. 

2.

 

Connect a digital voltmeter to the D1005. 

3.

 

Connect the voltmeter’s negative terminal to the 
black wire on the D1005 and connect the positive 
terminal to the red wire. The white wire is not 
used.

 

4.

 

If a D344-RL or D344-RT is installed, use the 
voltage monitor jacks as described below:

 

a.

 

The voltage measured by the voltmeter is half 
the sensitivity (in %/ft obscuration) of the 
D7050DH.  

b.

 

Multiply the voltage by 2. The result must be 
within the factory-marked calibration range 
printed on the label located on the bottom of 
the D7050DH. 

c.

 

If the D7050DH is outside the factory-marked 
calibration range, remove it and clean it. Or 
replace it as described in 

Section 7.1

 

Operational 

Testing

d.

 

After cleaning, recheck the calibration voltage 
measurement. If the D7050DH is still outside 
of the factory-marked calibration range, return 
it for recalibration to:  

National Repair Center 

130 Perinton Parkway 

Fairport, New York 14450 

8.0 Maintenance 

 

Notify all concerned parties before and 
after completing maintenance on or testing 
the  
fire alarm system. 

Clean the detector and base at least once each year 
using a vacuum or clean dry compressed air. Pay 
particular attention to the screens. You might need to 
clean more often in dusty areas or areas of heavy insect 
concentration. 
1.

 

Remove the D343’s cover. 

2.

 

Remove the D7050DH from the base and clean 
the base with a clean cloth and common window 
cleaner. 

3.

 

Remove the D7050DH’s cover using a thin,  
flat-blade screwdriver to pry the chamber from the 
cover. Insert the screwdriver into the cover slots 
and pry off the cover (

Figure 5

). 

Figure 5: 

Removing the 7050DH Cover 

1

3

4

2

2

2

 

1 -  Flat blade screwdriver tip inserted into a cover 

slot 

2 -  Cover slots (3) 

3 -  Chamber 

4 -  Cover 

4.

 

Remove the cover of the detector chamber. Gently 
pull the chamber cover up and away from the 
chamber (

Figure 6

Figure 6: 

Removing the D7050 Chamber Cover 

 

 

Содержание D7050DH

Страница 1: ...EN Installation Instructions Multiplex Photoelectric Duct Smoke Detector D7050DH ...

Страница 2: ...n The D7050DH Figure 1 is Underwriters Laboratories Inc UL Listed and requires a D343 Duct Detector Housing Use this detector with the D7024 or a DS9400M that includes a D7039 or DS9431 Multiplex Expansion Module An LED indicator on the head flashes every 3 to 8 sec to ensure the D7050DH has power and the smoke sampling circuitry functions This LED flashes at least once each second during an alarm...

Страница 3: ...ation and Installation Guide P N 31499 or the DS9400M Reference Guide P N 44578 for additional address limitations Figure 3 Setting the D7050DH Address 0 5 1 2 3 4 6 7 8 9 0 5 1 2 3 4 6 7 8 9 0 2 1 2 0 1 A B 1 2 3 1 Hundreds 2 Tens 3 Ones A D7024 FACP B DS9400 FACP For example 0 hundreds 9 tens 5 ones Address 95 on the D7024 FACP The A address range works on the D7039 Multiplex Expansion Module wi...

Страница 4: ...g to the next detector 7 2 Sensitivity Testing Calibration is important in determining a detector s continued operation Depending on local regulations calibration testing might be required more than once a year According to NFPA72 perform a Calibration Test at installation and every other year thereafter Perform a Functional Test monthly Test the D7050DH s sensitivity to meet NFPA 72 requirements ...

Страница 5: ...red by the voltmeter is half the sensitivity in ft obscuration of the D7050DH b Multiply the voltage by 2 The result must be within the factory marked calibration range printed on the label located on the bottom of the D7050DH c If the D7050DH is outside the factory marked calibration range remove it and clean it Or replace it as described in Section 7 1 Operational Testing d After cleaning rechec...

Страница 6: ...ion Refer to Section 7 0 Testing beginning on page 4 Do not paint the D7050DH Paint or other foreign matter covering the screens can stop or delay smoke from entering the detector 9 0 Specifications Table 3 column width table Standby Current Refer to the D343 Installation Instructions P N 48199 the DS9400M Reference Guide P N 44578 and the DS6431 Installation Instructions P N 41381 Alarm Current R...

Страница 7: ...D7050DH Installation Instructions Bosch Security Systems Inc 9 06 49325E 7 Notes ...

Страница 8: ... 2006 Bosch Security Systems Inc 49325E Bosch Security Systems Inc 130 Perinton Parkway Fairport NY 14450 9199 Customer Service 800 289 0096 Technical Support 888 886 6189 ...

Отзывы: