75 Individual PIN code
The Carephone 61 is delivered with the factory setting
246810
for the PIN
code. It is recommended not to change this code. If you need to change
this code, take care to write it down to find it easily. The PIN code is reset
when resetting the unit to its factory settings.
77 Pre-alarm time
The pre-alarm time of the Carephone 61 can be programmed. This is the
time within which an emergency call that has been initiated can still be
stopped.
–
Select a setting, in steps of 10 seconds, between
0
and
6.
–
0
= off,
1
= 10 s,
2
= 20 s, etc.
81-89 & 80 Assigning wireless transmitters
The registered wireless transmitters can be programmed to trigger
different alarm functions, e.g. a fire alarm. First, register the wireless
transmitters on the Carephone 61; refer to programming steps 90-99. To
assign a function to wireless transmitter 91, select programming step 81.
–
0
= emergency call with a wireless transmitter (see step
50
)
–
1
= emergency call with the Carephone 61 (see step
51
)
–
4
= action button (see step
71
)
–
5
= external input (when step
73
is set to
3
)
–
7
= connection to a destination number (with acknowledgement)
–
8
= output assigned in programming step
72
91-99 & 90 Registering wireless transmitters
Up to 10 wireless transmitters or detectors can be registered with a code
number contained in the transmitter itself. These code numbers must be
stored in the Carephone 61. To register a wireless transmitter:
a.
Select programming step 91.
b.
The Carephone 61 announces:
"Selection 91 is..."
c.
Press the button of the corresponding wireless transmitter.
d.
The Carephone 61 announces:
"Selection 91 is..."
e.
Press
C
to confirm.
The wireless transmitter is now stored. Restoring the factory settings will