Appendix G — Calibration
The
GasFinder2
is shipped already calibrated and does not require any calibration in the fi eld.
The calibration of the
GasFinder2
is done
by passing a known concentration of gas
through a test cell, which is placed in the
path of the laser beam.
The calibration data are stored in the
instrument’s software as a standard
reference waveform.
During operation, the
GasFinder2’s
internal reference cell is compared with this stored
waveform at frequent intervals. Any signifi cant deviation generates a status code to alert the
user to a potential calibration problem. The
GasFinder2
software can be used to check the
reference cell, as well as to download sample, reference, and calibration waveforms (System
Menus/Internal Arrays Transfer) to verify that the internal calibration system is functioning
correctly. If
GasViewOP
is being used, these waveforms can be displayed on the computer
(Port1/Arrays Plot Enable and press the ‘Arrays Transfer’ in the
GasFinder2
menu)
.
A technical note detailing the calibration procedure and quality assurance is available from
Boreal Laser Inc.
Содержание GasFinder2
Страница 52: ...9 2 Boreal Laser Inc GasFinder2 Operation Manual Appendices July 17th 2017 This page left intentionally blank...
Страница 66: ...9 16 Boreal Laser Inc GasFinder2 Operation Manual Appendices July 17th 2017...
Страница 68: ...10 2 Boreal Laser Inc GasFinder2 Operation Manual Drawings July 17th 2017 This page intentionally left blank...
Страница 69: ......
Страница 70: ...10 2 Boreal Laser Inc GasFinder2 Operation Manual Drawings August 2011 This page intentionally left blank...
Страница 71: ......
Страница 72: ...10 2 Boreal Laser Inc GasFinder2 Operation Manual Drawings August 2011 This page intentionally left blank...
Страница 73: ......
Страница 74: ...10 2 Boreal Laser Inc GasFinder2 Operation Manual Drawings August 2011 This page intentionally left blank...