Integrated Multi-parameter Sensor Operation manual
84
2564
temperature of the fourth
point standard solution
calibration of electrode 4
float
read
2566
fifth point calibration of
electrode 4
float
write
input the standard solution
value
2568
the fifth point standard
solution value of
electrode 4
float
read
2570
electric voltage of the
fifth point standard
solution calibration of
electrode 4
float
read
2572
temperature of the fifth
point standard solution
calibration of electrode 4
float
read
Note: electrode 1-4 channels, customers can choose to access, according to the access channel
corresponding to the corresponding address.
Chapter 9 Problems and Solutions
phenomenon
Possible factors
Method of exclusion
Conductivity
Display
instability
A, there are bubbles in the
pipeline
A, rectification of the pipeline or
alternative measurement points
Large reading
B, unstable water quality
B. Reasons for using a stable water
source to remove the meter
Dissolved
Low test
Inappropriate flow rate
Install the conductivity cell at a