Integrated Multi-parameter Sensor Operation manual
17
Figure 14 Fluoride ion sensor
Fluoride sensor Specification
principle
ion-selective electrode method
Range
0.5~1000 mg/L
resolution
0.01~1 mg/L depending on the range
Accuracy
±
10 % or
±
2 mg/L,
(the electrode is greatly affected by
temperature, field comparison is the best field calibration,
otherwise it may be out of range)
2.13 Chloride Ion Sensor
CL-ion selective electrode is a kind of solid membrane electrode. A high strength AgCl/Ag 2S compact
solids is the main material of the membrane, which then is sealed in the plastic with resin. Sliver is
adopted as the conductive electrode for the surface of solid. The appearance of chloride ion sensor is
showed by the Figure 15
Figure 15 Chloride ion sensor