Integrated Multi-parameter Sensor Operation manual
57
6.2.6 Turbidity calibration
Get into the interface of turbidity calibration, and press the key of “OK” at the site of “wiping before
calibration”. Press the key of “Enter” to perform cleaning procedure for one time by scratching, and
eliminate the influence of bubbles on turbidity calibration process.
Choose first point/second point turbidity calibration according to actual requirement to make integrated
multi-parameter analyzer immerge into standard solution, and at the same time, input corresponding
standard solution value in the square frame of the hand manipulator. Then, choose the key of “OK” and
press "Enter" for calibration and confirm the calibration status according to the prompt of calibration
status. When the calibration is indicated to be successful, the calibration at this point is complete.
6.2.7 Chlorophyll calibration
Get into the interface of chlorophyll calibration, and make integrated multi-parameter analyzer immerge
into distilled water or rinsing. Press the key of “OK” at the site of “wiping before calibration”. Press the
key of “Enter” to perform cleaning procedure for one time by scratching, and eliminate the influence of
bubbles on chlorophyll calibration process. Press the key of “OK” at the site of “zero calibration” in the
interface, and press the key of “Enter”.