
Integrated Multi-parameter Sensor Operation manual
54
Pour the standard solution at point 2 (generally recommended 1 mg/L) into a clean, dry or pre-rinsed
calibration container. Carefully immerse the probe at the end of the main unit in the solution. Please
ensure that the front end of the Chloride Ion Sensor is immersed in the solution at least 2 cm and enter the
standard value. When the value in the software interface is stable, click “2 point Calibration”. Rinse and
dry the sensor and calibration container with water after completion.
Pour the standard solution at point 3 (generally recommended 200 mg/L) into a clean, dry or pre-rinsed
calibration container. Carefully immerse the probe at the end of the main unit in the solution. Please
ensure that the front end of the Chloride Ion Sensor is immersed in the solution at least 2 cm and enter the
standard value. When the value in the software interface is stable, click “3 point Calibration”. Rinse and
dry the sensor and calibration container with water after completion.
You can directly select the input value in the correction parameter box and click “factor” and “deviation”
to calibrate the factor and deviation value.
6.2 Calibrate the sensor by hand manipulator
Choose “calibration” in the main menu page, and press the key of “Enter” to confirm. Then, you can get
into the calibration interface of selective parameters. The parameters you choose must be the probes that
have been connected as showed in the diagram below.
Please check the details of all the specific calibration steps of sensors below
in the Chapter 6.1