11

kHz and GHz Keys

The kHz and GHz keys are provided for convenience when entering 
frequency; however, the display will only indicate in MHz. Similarly, 
the V key can be used for entering input level; however, the display 
will indicate in millivolts.

ENTER key

The ENTER key is used for unitless quantities, such as special function 
and program numbers.

CLR key

The CLR key is used to recover from errors. Without changing func-
tion, depress the dBm key. The FREQUENCY/LEVEL display will now 
indicate Error 9 or 11. This means that an inconsistent units key has 
been depressed to terminate a data entry.

Depress the CLR key. The display will return to normal. A list of error 
codes is presented in Table 3-8, at the end of the section 3-26 of the 
8201A instruction manual.

LEVEL Function key

Depress the carrier LEVEL function key, then the 0 and dBm keys. The 
FREQUENCY/LEVEL display will now indicate 0.00 dBm.

mV/MHz key

Depress the mV/MHz key. The display will change to 223.6 m V. 
Carrier level may be entered in millivolts, volts, or dBm. The control 
program will recalculate or rescale numbers as required.

The DEL and CLR keys are also labeled as up- and down-arrow keys 
for scrolling through SPCL function menus.

Getting familiar with Measurement 
Control Keys

The measurement control keys consist of the groups of switches 
marked FILTERS and DETECTORS. These keys may be operated at any 
time and will affect the MODULATION and AUDIO displays. The filter 
switches are arranged as self-cancelling groups of four and five keys. 
Depressing any high-pass switch will cause the selected filter to be 
placed into the measurement channel and cancel any other selected 
high-pass filter. Similarly, depressing any low-pass filter will cause 
the selected filter to be placed into the measurement channel and 
cancel any other selected low-pass filter. 

Figure-6: Measurement Control Keys of 8201A Modulation Analyzer

HIGH-PASS Filter keys

Depressing any high-pass switch will cause the selected filter to be 
placed into the measurement channel and cancel any other selected 
high-pass filter. 

LOW-PASS Filter keys

Similarly, depressing any low-pass filter will cause the selected filter 
to be placed into the measurement channel and cancel any other 
selected low-pass filter.

The maximum low-pass bandwidth selection is a function of carrier 
frequency. The control program will automatically adjust the low-
pass filter cutoff frequency as required. The carrier breakpoints and 
low-pass filters are:

Carrier 

 Filter 

< 500 kHz 

15 kHz max 

< 10 MHz  

50 kHz max 

> 10 MHz  

220 kHz max

DE-EMPHASIS Filter Keys

The de-emphasis filters are normally available when measuring FM 
only. They are selected in the same manner as the high-pass and 
low-pass filters, but are automatically removed from the measure-
ment channel when AM or PM modulation function is selected. The 
selected de-emphasis filter will be restored when the FM function is 
again selected.

Additionally, the de-emphasis filters may be placed before or after 
the modulation display. This is accomplished by selecting SPCL func-
tion 7 for pre-display and SPCL function 8 for post-display de-em-
phasis. SPCL function 9 permits the de-emphasis filters to be selected 
in the AM measurement function. This is useful for performance 
verification of the filter 3 dB points.

Содержание 8201A

Страница 1: ...Taking performance to a new peak Quick Start Guide 8201A Modulation Analyzer PN 98407700A ...

Страница 2: ...ent to use This document briefly describes the operat ing information for the Model 8201A Modulation Analyzer and all 8201A S variants including the descriptions of the front and rear panel controls displays connectors option selections and instructions for local and remote modes of operation Typical measurement situations are also described The 8201A is an updated version of the discontinued Mode...

Страница 3: ...Displays and Operating Controls 5 Getting Started 5 Getting familiar with Function Keys 10 Getting familiar with Data Keypad 10 Getting familiar with Measurement Control Keys 11 Other Keys 12 Displayed Messages 13 Special Functions 13 Calibration and Performance Tests 14 ...

Страница 4: ...IMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FIT NESS FOR A PARTICULAR PURPOSE Boonton will not be liable for any incidental damages or for any consequential damages as defined in Section 2 715 of the Uniform Commercial Code in connection with products covered by the foregoing warranties Safety and Caution Summary The following general safety precautions must be observed during all phas...

Страница 5: ... listed in Table 1 Table 2 and shown in Figure 2 and Figure 3 Figure 1 8201A Modulation Analyzer Displays Getting Started The following instructions walk through a basic operation and func tionality tests to verify that the unit is working properly 1 Turn ON the instrument and press LCL INIT key 2 After a while the FREQUENCY LEVEL display will show up with 8201A firmware reference number and the o...

Страница 6: ...nd it displays modulation in for AM kHz for FM deviation and RAD for PM deviation Also displays ratio in or dB BUS Status display 4 Displays current IEEE 488 bus status REM remote enabled LSN listener ad dressed TLK talker addressed and SRQ service request active SPCL PRGM display 5 The SPCL PRGM display is an entry only display which is two characters wide and display the current program number o...

Страница 7: ...andpass filter Selects the 30 Hz high pass or CCIR bandpass filter Selects the 300 Hz high pass or CCIIT bandpass filter Selects the 3000 Hz high pass or C MSG bandpass filter Toggles filter selection from high pass to bandpass Low pass Hz keys 3 15 20 50 220 11 Selects the 3 kHz low pass filter Selects the 15 kHz low pass filter Selects the 20 kHz low pass filter Selects the 50 kHz low pass filte...

Страница 8: ...fore selecting program number for store or recall LCL INIT key 22 Causes instrument to go to Iocal if local lockout is not active and remote is ac tive or initialize when the instrument is in the local state STO key 23 Stores the instrument setup at selected program number RCL key 24 Recalls instrument setup from selected program number PEAK keys key key key 25 Selects peak detector for display Se...

Страница 9: ...nal to external test equipment AM out connector 7 Provides a means for connecting the demodulated AM signal to external test equipment DIST out connector 8 Provides a means for connecting the distortion analyzer signal output to external test equipment EXT REF connector 9 Provides a means for connecting an external 10 MHz frequency standard to the internal time base circuits Optional CAL OUT conne...

Страница 10: ...described as below The top row of illuminated switches are the function keys These keys are used to select the parameter to be displayed and to enable the data keypad for subsequent data entry The functions are carrier FREQ carrier LEVEL audio SINAD audio FREQ audio DIST AM FM PM SPCL and PRGM The LED in the switch of the selected function will be illuminated continuously the others will be off un...

Страница 11: ...switch will cause the selected filter to be placed into the measurement channel and cancel any other selected high pass filter Similarly depressing any low pass filter will cause the selected filter to be placed into the measurement channel and cancel any other selected low pass filter Figure 6 Measurement Control Keys of 8201A Modulation Analyzer HIGH PASS Filter keys Depressing any high pass swi...

Страница 12: ... be used with any detector In operation as modulation measurements are made the larger of the current measurement or the previous measurement becomes the displayed modulation Depress the HOLD key to activate this mode and then depress the key again to cancel Other Keys Figure 7 Other Keys of 8201A Modulation Analyzer RATIO Key The RATIO key is an alternate action key which changes the active displ...

Страница 13: ...ake about 80 seconds to complete The AM detector is cali brated first followed by the rms detector the FM detector and finally the PM detector If calibration error occurs they will be displayed as the particular is being calibrated Displayed Messages UNLOC message When the Model 8201A unlocks the FREQUENCY LEVEL display will be overwritten with the UNLOC message and the AUDIO and MODU LATION displ...

Страница 14: ...y PM 1 0 accuracy FM Calibration The calibration process consists of 1 Applying to the FM discriminator in alternation two accurately controlled frequencies 2 Measuring the resulting recovered modulation information 3 Computing a correction factor for subsequent FM measure ments AM Calibration The operation of the AM calibrator is similar to that of the FM calibrator For detailed theory of operati...