- 11 -
Micro-motor
Remove all attachments to the micro-motor prior to cleaning. The outer surface of the
micro-motor should be cleaned with an antiseptic solution.
Never oil the micro-motor.
Make sure no moisture gets inside the micro-motor.
Warning
:
Do not dip the micro-motor and extension wire directly into the sterilization
chemical. The sterilization chemical will interfere with normal operations if
it reaches the inside of the system.
Warning:
Do not leave the disinfectant on parts longer than recommended by the
disinfectant manufacturer to avoid damaging materials.
Note: The hand piece should be cleaned after each use between patients.
Straight Hand Piece Attachment (optional accessory)
The straight hand piece should be lubricated at the end of each day of use. Drop 2-3 drops of
mineral oil into the top and allow it to flow down through the hand piece.
Prophy Head (optional accessory)
The prophy angle should be cleaned to remove hair and debris after each patient. At the end of
the day it should be cleaned and lubricated. To clean the Prophy Head, follow the steps
below.
1. Remove the rubber cup
2. Remove the cup holder with pliers by turning the knurled knob right (the knurled knob has a
reversed thread). Wash away debris from the inside. Then put 1 drop of oil onto the gears
(3-in-1 multipurpose oil works well).
3. Make sure all hair and fur, if any, is removed from the end of the prophy angle.
4. Replace the knurled knob.
Section V: Installation
5-A: General Information
If the installation of your KRUUSE-SP1 system is performed by non-Bonart distributor personnel,
check to see that the requirements below are followed.
5-B: Water Line Requirements
Incoming water supply line pressure to the Scaler must be 25 psi (172 kPa minimum) to 60
psi (414Kpa maximum). If your dental water supply line pressure is above 60 psi, install a
water regulator on the water supply line to your KRUUSE-SP1 unit.
After the above installation requirement is fulfilled, thoroughly flush the water prior to
connecting to the scaler.