Conditions for Operation
This equipment is for indoor use only.
Environment temperature: 10-40
℃
.
Relative humidity: 10-70% (Non condensing).
The accuracy of the torque, rotation speed and apical position is guaranteed only when the
original contra angle is used which is supplied by manufacturer.
Do not sterilize the central unit, nor place it into autoclave or ultrasonic tank.
Do not use the system in the presence of free oxygen or flammable gas mixtures.
Portable and mobile RF communications equipment can affect Medical Electrical equipment.
Do not use RF equipment outskirts for the product.
Follow the instruction of file manufacture to set the rotation and speed of the motor.
If the Central unit has not been used for long time, check it before use
There is electric circuit which control the torque (TORQUE LIMITER Function) to prevent
files from breaking; however, files may still break due to metal fatigue if the torque is
conditioned to be higher. Please check the working instruction of file before use.
The battery charger must be supplied at a voltage in the range: 100V-240V (+/-10%), 50-60Hz
(+/-10%). Use only original parts.
Should any anomalies arise during operation, suspend work and contact your technical service
center.