Replacement of Fuses
There are two supply fuses located in the fuse drawer. To replace the fuses:
•
Disconnect the unit from the power supply.
•
Remove the plug from the socket in the back of the unit.
•
Pull back on the fuse drawer (see Figure 11).
•
Pull out the fuse drawer.
•
Check and replace with the correct fuses if necessary. The fuses must be
5mm x 20mm quick acting, rated 250V.
Model 241000:
-2AF
Model 241000-2:
-1AF
•
Push the fuse drawer back in. Reconnect unit to the power supply.
9. Warranty
When used in laboratory conditions and according to these operating instructions, Boekel
warrants this product to be free of defective material and workmanship for a period of two
years from the date of manufacture. The liability of Boekel for any defective equipment
during the warranty period shall be limited to the repair of such equipment or replacement
thereof without charge for parts or labor.
10. Service
A Returned Goods Authorization (RGA) number must be obtained before any Boekel
products are returned for any reason. A Decontamination Notice must be completed,
signed by the user, and returned to Boekel Scientific prior to receiving the RGA number.
Please be sure to mark the outside of the return goods package with this RGA number to
ensure prompt handling.