8.2

Replacement of Fuses 

There are two supply fuses located in the fuse drawer.  To replace the fuses: 

Disconnect the unit from the power supply. 

Remove the plug from the socket in the back of the unit. 

Pull back on the fuse drawer (see Figure 11). 

Pull out the fuse drawer.   

Check and replace with the correct fuses if necessary.  The fuses must be 
5mm x 20mm quick acting, rated 250V. 

                           Model 241000: 

-2AF 

   

 

Model 241000-2: 

-1AF 

 

Push the fuse drawer back in.  Reconnect unit to the power supply. 

9.     Warranty 

When  used  in  laboratory  conditions and  according to  these  operating  instructions,  Boekel 
warrants this product to be free of defective material and workmanship for a period of two 
years  from  the  date  of  manufacture.    The  liability  of  Boekel  for  any  defective  equipment 
during the warranty period shall be limited to the repair of such equipment or replacement 
thereof without charge for parts or labor. 

10.   Service 

A  Returned  Goods  Authorization  (RGA)  number  must  be  obtained  before  any  Boekel 
products  are  returned  for  any  reason.    A  Decontamination  Notice  must  be  completed, 
signed  by  the  user,  and  returned  to  Boekel Scientific  prior  to  receiving  the  RGA  number. 
Please  be  sure  to  mark  the  outside of the  return goods package  with  this RGA  number  to 
ensure prompt handling. 

Figure 11

Fuse
Drawer

Содержание InSlide Out 241000

Страница 1: ...N2400117 Rev 4 01 2013 Boekel InSlide Out Hybridization Oven Models 241000 and 241000 2 Operating Instructions ...

Страница 2: ...Lamps see Figure 1 3 4 2 Setting the Temperature 4 4 3 Loading Tray with Slides 4 4 4 Loading Tray into Heating Chamber 4 4 5 Removing Tray from Heating Chamber 5 5 Accessories 5 5 1 Slide Holder 5 6 Fault Diagnosis 6 7 Technical Specifications 6 8 Maintenance and Service 7 8 1 Cleaning 7 8 2 Replacement of Fuses 7 9 Warranty 8 10 Service 8 ...

Страница 3: ...ntrinsic protection may be impaired Allow unit to dry out for 4 hours minimum Connect only to a power supply with a voltage corresponding to that on the serial number label Connect only to a power supply that provides a Protective Earth terminal Before moving disconnect at the power supply socket Do not remove the plug Do not check the temperature by touch but instead use the temperature display T...

Страница 4: ...ain for future shipment or storage of the unit Inspect for damage Report all shipping damage to the carrier immediately Shipping damage is covered by the carrier and repair replacement for shipping damages must be coordinated through the carrier Complete and return the Warranty Registration Card Packs should contain In Slide Out Hybridization Oven Power Line Cord Operating Instructions Tray Assemb...

Страница 5: ...vent of power loss the Temperature Controller retains the last set temperature value 4 3 Loading Tray with Slides The tray with the standard wire rack can be loaded with up to 20 standard glass slides see Figure 2 Using the optional slide holder up to 18 standard glass slides can be loaded see Figure 3 Use your standard protocol to prepare the slides with probe and buffer Place filter paper at the...

Страница 6: ...the main door 4 5 Removing Tray from Heating Chamber Extreme care must be used when removing the tray due to the potentially high temperature It may be necessary to use gloves when removing the tray To remove the tray from the Heating Chamber open the main door lift the Latch Handle up and rotate it 180 in a counter clockwise direction Rotate the latch down lift the frame up and carefully slide th...

Страница 7: ...re is lower than set temperature b Temperature control circuit fault a Check set temperature b Have unit checked by competent person 4 Slides dry out during heating cycle a Gasket not sealing b Lid bent c Insufficient buffer placed in bottom of tray d Latch not fully engaged a Replace Tray and Gasket Assembly b Replace Lid c Place more buffer in bottom of tray d Re install Tray Assembly into chamb...

Страница 8: ...Service All Boekel laboratory products are designed to comply with IEC1010 1 No routine maintenance is required 8 1 Cleaning Disengage power cord prior to cleaning If a spill occurs use appropriate clean up procedures as required for radiation or biohazard control The outer casing may be cleaned with water and a damp cloth Do not submerge or immerse the InSlide Out in water Before using any cleani...

Страница 9: ...to these operating instructions Boekel warrants this product to be free of defective material and workmanship for a period of two years from the date of manufacture The liability of Boekel for any defective equipment during the warranty period shall be limited to the repair of such equipment or replacement thereof without charge for parts or labor 10 Service A Returned Goods Authorization RGA numb...

Отзывы: