7. Service mode and adjustment
Place the first adjustment strip in the middle of the insertion
area and initiate the adjustment procedure by pressing
<Start>.
The adjustment of the first adjustment strip begins. Once
adjustment is completed, a report will be printed and sent to
the printer interface. If the adjustment was carried out
correctly the display reads:
Place the second adjustment strip in the middle of the
insertion area and initiate the adjustment procedure by
pressing <Start>.
The adjustment of the second adjustment strip begins.
Once adjustment is completed, a report will be printed and
sent to the printer interface. If the adjustment was carried
out correctly the display reads:
After the adjustment has been completed, the Service
Mode is terminated and the analyzer returns to the Normal
Mode.
"Insert second adjustment strip"
Insert ADJ.STRIP
"Analyzer is in the Normal
Mode"
READY <START>