Service Manual Miditron
®
Junior
II
/ ID 1997491 / MJ / 1.0 - June 1998 / Page 30
Cal. Error 50 - 85
Calibration not valid.
Recalibration necessary.
Cal. Error 10 - 45
Calibration not valid.
Measurement of a second
calibration strip necessary
Display message:
"Recalibrate !"
Difference of < 1 % R between 1st and 2nd
calibration strip
Cal. Error 10 - 45
Calibration not valid.
Recalibration necessary.
Stored
calibration is still
valid.
Calibration is valid.
New calibration values
will be stored.
Difference of < 10 % between measured value and reference value
Difference of < 1 % R
Yes
No
No
Yes
No
Yes
5. Calibration
Miditron
®
Junior
II
is ready to read as long as there are
valid calibration values stored in the analyzer. Calibration
every 2 weeks is recommended. If calibration cannot be
carried out for any reason, e.g. because there are no
Control-Test M calibration strips available, you can bypass
the calibration procedure and continue reading by pressing
<Start> at the "Please Calibrate !" prompt. Miditron
®
Junior
II
then uses the stored calibration values to carry out further
readings.
Note: In this case, "Calibration is necessary" will be printed
on the results report.
Boehringer Mannheim can give no guarantee as
to the accuracy of results if the analyzer has not
been calibrated.
Содержание Miditron Junior II
Страница 1: ...Service Manual Miditron Junior II ID 1997491 MJ 1 0 June 1998 Page 1 Miditron Junior II Service Manual...
Страница 73: ...Service Manual Miditron Junior II ID 1997491 MJ 1 0 June 1998 Page 73 10 Electronics Sheet2 PCB Main Part 2...
Страница 74: ...Service Manual Miditron Junior II ID 1997491 MJ 1 0 June 1998 Page 74 10 Electronics Sheet3 PCB Main Part 3...
Страница 75: ...Service Manual Miditron Junior II ID 1997491 MJ 1 0 June 1998 Page 75 10 Electronics PCB Main Part 4 Sheet4...
Страница 132: ...Service Manual Miditron Junior II ID 1997491 MJ 1 0 June 1998 Page 132 14 Spare Parts...