Guidelines and manufacturer’s declaration – electromagnetic immunity
The PressureTel
™
blood pressure monitors are intended for use in the electromagnetic environment specified below. The customer or the user of the PressureTel
™
blood pressure monitor should ensure that it is used
in such an environment.
Immunity test
IEC 60601-1 test level
Compliance level
Electromagnetic environment – guidelines
Electrostatic discharge (ESD)
+- 6 kV contact discharge
+- 6 kV contact discharge
Floors should be wood, concrete or ceramic tile. If floors are
IEC 61000-4-2
+- 8 kV discharge in air
+- 8 kV discharge in air
covered with synthetic material, the relative humidity should
be
at
least
30%.
Power frequency (50/60 Hz) magnetic field
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
IEC 61000-4-8
characteristic of a typical location in a typical commercial and
hospital
environment.
10 Manufacturer’s electromagnetic guidelines